Deckma Hamburg OMD-2008 Instruction Manual

15ppm bilge alarm
Hide thumbs Also See for OMD-2008:

Advertisement

I N S T R U C T I O N M A N U A L
15ppm Bilge Alarm
OMD-2008
Standard Version
and Versions with
Electric Valve (EV)
Flow Control (FC)
(MCU)
Manual Cleaning Unit
DECKMA HAMBURG GmbH
Kieler Straße 316, D-22525 Hamburg - Germany
Tel.: +49 (0) 40 54 88 76-0, Fax: +49 (0) 40 54 88 76-10
Internet: www.deckma.com
eMail: post@deckma.com

Advertisement

Table of Contents
loading

Summary of Contents for Deckma Hamburg OMD-2008

  • Page 1 Versions with Electric Valve (EV) Flow Control (FC) (MCU) Manual Cleaning Unit DECKMA HAMBURG GmbH Kieler Straße 316, D-22525 Hamburg - Germany Tel.: +49 (0) 40 54 88 76-0, Fax: +49 (0) 40 54 88 76-10 Internet: www.deckma.com eMail: post@deckma.com...
  • Page 2 Alternatively the unit may be replaced by a calibrated 15 ppm Bilge Alarm. The OMD-2008 EVFC is designed in that way, that only the measuring cell needs to be changed, as this unit carries the Calibration Certificate. The Calibration Certificate with the date of the last calibration check should be retained onboard for inspection purposes.
  • Page 3 The purchaser shall bear the cost and risk of transport of defective parts and repaired parts supplied in replacement of such defective parts. ANY DISMANTLING OR BREAKING OF A SEAL WILL VOID THE WARRANTY Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 2 of 28...
  • Page 4: Table Of Contents

    Calibration and repeatability check ................37 14.2 Function Test at Classification Survey and Port State Control ......... 38 Spare Parts......................39 15.1 Recommended On Board Spares ................39 REMARKS ......................40 Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 3 of 28...
  • Page 5: Introduction

    1.2 Instrument Versions and Options The OMD-2008 is available in different versions. It may be equipped with an electric switchover valve for clean water usage, and a Flow Sensor. A Manual Cleaning Unit (MCU) is available as an option. Different instrument versions are covered in this document.
  • Page 6 DECKMA HAMBURG GmbH 1.2.2 EV (Electric clean water Valve) OMD-2008 EV instruments are equipped with an electric switchover valve for the clean water stream. The valve can be controlled via the menus, or remotely using the control input. 1.2.3 MCU (Manual Cleaning Unit) Optionally the instruments can be fitted with a Manual Cleaning Unit.
  • Page 7: Important Notes

    In case of freezing conditions the measuring cell should be emptied completely. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 6 of 28...
  • Page 8: Principle Of Operation

    Section 10.4. In this mode also the individual adjustment of the time delays for the alarms can be done. Both alarm circuits are also related to an alarm LED on the front panel. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 7 of 28...
  • Page 9: Electronic Record

    SYSTEMFAULT relay) will switch to alarm condition. 3.4 Electronic record The OMD-2008 keeps an electronic record. The recorded data can be viewed at the instrument at any time. Browsing the data does not interfere with the measurement or instrument operation. Data is stored in excess of 18 Months.
  • Page 10: Specifications

    + 1 to + 55° C Sample Water + 1 to + 65° C Temperature: Roll: Up to 45° Degree of Protection: IP 65 Pipe Connections: for 6mm OD Cu or SS pipe Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 9 of 28...
  • Page 11 200 mm W x 200 mm H x 100 mm D Size (EV-FC-Arrangement) 150 mm W x 290 mm H x 140 mm D Weight: 5,9 kg Technical specifications are subject to change without notification Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 10 of 28...
  • Page 12: Construction

    DECKMA HAMBURG GmbH 5 Construction 5.1 Overview OMD-2008 instruments consist of two main components: The Computer Unit contains the display PCB with the data logger and the terminals for external connections. The Measuring Cell assembly consists of the Measuring Cell, a valve to select sample water stream, or clean water stream, and, depending on the version of the instrument, additional devices like a flow sensor.
  • Page 13: Other Devices

    5.4 FW valve The OMD-2008 uses a switchover valve to control the usage of clean water for cleaning, or zeroing purposes. Whenever the FW valve is switched to Clean water position, and not to sample stream position (normal operating position), the instrument will be in alarm condition.
  • Page 14 DECKMA HAMBURG GmbH Computer Unit Flow Sensor Measuring Cell Clean Water Valve EV Manual Cleaning Unit (MCU) (Option) Fig. 1 Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 13 of 28...
  • Page 15: Installation

    A sample stream flow rate of approx. 2 Liters per Minute has to be provided. Mount the OMD-2008 Computer Unit by means of M6 or M8 screws on to a rigid vertical surface and preferably with the display panel of the monitor at eye level.
  • Page 16 Care must be taken at mounting of the pipes connections to avoid any torsion of the housing and damage of the instrument. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 15 of 28...
  • Page 17: Piping

    The sample stream through the instrument is recommended to be approx. 2 Liters per Minute. Connect the OMD-2008 EV-FC Monitor to the sample point of the oily-water separator outlet and to a source of oil free water. Fittings on the Instrument are for 6 mm OD copper or stainless steel pipe.
  • Page 18 DECKMA HAMBURG GmbH Fig. 3 Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 17 of 28...
  • Page 19: Wiring

    (PE bolt inside, right side). The instrument provides a pilot voltage output at Terminals 3&4. This is internally connected to the power supply input (Terminals 1&2 via Fuse T2A). The pilot Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 18 of 28...
  • Page 20 Alarm contacts description is in alarm (non-energized) condition. Note: In case the OMD-2008 EV-FC is installed as replacement to a MEPC.60(33) approved separator and no “Status Separator” signal is available, a link across terminal 11 + 12 can be installed. Alternatively a “Status Separator”...
  • Page 21: Typical Control System

    The unit is designed for a power supply of 24 V to 240 V AC or DC. It has an automatic voltage selection. The power supply must have a fuse rated no more then 2A. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 20 of 28...
  • Page 22: Commissioning

    Minute. Flow rate for both sample stream and clean water stream should be approx. 2 Liters per Minute. On OMD-2008 FC instruments the flow rate can be read from the display.
  • Page 23: Functional Tests

    If the sample flow rate is too small, the instrument will be in Alarm and display “Status: Flow”. On instruments fitted with Manual Cleaning Unit (OMD-2008 EV-FC MCU), sample a higher or a lower flow rate can be adjusted. To get the higher flow rate, remove the smaller O-Ring (9.5x2) from the Manual Cleaning Unit.
  • Page 24 If the flow rate is not sufficient, the instrument will display “Status: Flow”. Both sample flow rate and clean water flow rate are recommended to be approx. 2 Liters per Minute. g) Switch the electric Valve (EV) back to sample stream. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 23 of 28...
  • Page 25: Programming Mode

    Refer to Fault Pressing the finding table in button will display manual for additional explanations of information. status information. Pressing the button will display more detailed Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 24 of 28...
  • Page 26 “+” button. factory default Confirm with “OK” values. To reset to to reset all settings factory values once to the factory again press the default settings. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 25 of 28...
  • Page 27 Confirm with “OK” additional menu. to activate the information should be displayed. Flow rate and status of the meter is displayed. Exit from Flow display by pressing the ESC button. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 26 of 28...
  • Page 28 With the data. With the invoked. button it can be button it can be In both data display modes the arrow buttons can be used to navigate to another date/time Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 27 of 28...
  • Page 29 DECKMA HAMBURG GmbH NB: Changed values have to be confirmed by pressing the " OK " button. Otherwise the existing values remain valid. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 28 of 28...
  • Page 30: Operating Instructions

    Whenever clean water is used, the instrument will be in Alarm condition. The OMD-2008 EV-FC has an additional Flow Sensor to detect the flow rate of the water through the instrument. If the sample does not flow, the instrument will be in alarm condition.
  • Page 31: Operator Maintenance

    Please note that a desiccator package should never be opened unless it is installed immediately. j) Switch Clean Water Valve to SAMPLE setting. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 30 of 28...
  • Page 32: Manual Cell Clean Unit (Mcu)

    Check the reading is between 0 to 2 ppm. d) Repeat a) to c) at least once a week or as necessary. e) Additional conventional cleaning with the cleaning brush may be necessary. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 31 of 28...
  • Page 33 NB: The Manual Cell Clean Unit may also be used during normal operation with sample water, but in this case an alarm occurs because the wiper is passing the light source. MCU Spares: Wiper Seal DH 77606 Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 32 of 28...
  • Page 34 Just cleaning with brush and clean water will not help in this case. 12.5 Cleaning of Glass Tube at 15 ppm Bilge Alarms OMD-2008 series IMPORTANT: NEVER DISASSEMBLE THE UNITS AS THIS MAY VOID THE...
  • Page 35: Fault Finding

    13 Fault finding See Section 2 for important notes. The OMD-2008 EV-FC will indicate several malfunctions in the status line of the display. Pressing the “OK” button will lead into an information window, similar to the items listed in the table below.
  • Page 36 Gree Alarm Check Sample sample flow Blinki flow rate rate Flow Input (Terminals 15&16) open Red / Alarm Check Stea communic connection ation between between computer unit computer Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 35 of 28...
  • Page 37 Blinki operati humidity critically high (>40%rH) Gree Norma Sample Check/Repla temperatur ce Desiccator Blinki operati below dewpoint. Instantane condensati possible Red / Alarm Internal Restart the Stea error system Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 36 of 28...
  • Page 38: Memory Card

    Under normal use the card should not be taken out, as this is linked with the specific system. The card can be read in other OMD-2008 units, but writing is only possible in the related system.
  • Page 39: Function Test At Classification Survey And Port State Control

    Note that this test will only switch the alarm contacts to non-alarm condition, if the sample contains less than 15ppm oil content and all other conditions for proper measurement are OK. Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 38 of 28...
  • Page 40: Spare Parts

    Wiper Seal for Cell Clean Unit 77606 OMD-CR Memory Card Reader 75185 Function Test Kit 18510 Flow Sensor 16700 15.1 Recommended On Board Spares DESCRIPTION ART-NUMBER 1 off Service Kit 18570 Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 39 of 28...
  • Page 41: Remarks

    All the modifications and deviations from the standard form, which have to be carried out in the supply, should be attached at this paragraph. Commissioned on: ......by: .......... Date Firm's Name Remarks: Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions) Page 40 of 28...