1 .2 .7
Programmability
Frequency is
programmable via
rear panel
input connector using BCD-coded TTL voltages
or BCD-coded contact closures .
1 .2 .8
General
OPERATING TEMPERATURE
25 t5oC, all specifications apply
25 t15 0 C, with slight degradation of speci-
fications
OUTPUT CONNECTOR
Type N
RFI
<1
VV is induced in a two-turn, one-inch
diameter loop which is held one inch away
from any surface .
Loop feeds a 50 ohm re-
ceiver ;
POWER
115/230 V ±10%, 50/60 Hz, 40 VA
DIMENSIONS
12 in .
(30 .3 cm) wide,
5k in .
(13 .4 cm)
high, 13 3/4 in . (34 .9 am) long
WEIGHT
25 lb .
(11 .4 kg) net,
30 lb . (13 .6
kg)
shipping .
1 .3
OPTIONS
Options 1A, 1B, 4, and 7 are factory installed .
Option 3 may be factory or field installed .
1 .3 :1
RF Level Programming
For both Options 1A and 1B, the instruments
are calibrated for +13 dBm at 50 MHz
like
a standard unit, but due to greater losses
in
Programmable Attenuators, a calibrated
output is only guaranteed to +12 dBm.
Option lA Program Level Range :
0 to 109 .9
dB
in
.1
dB
steps
(programmed
via
rear-panel plug) .
0
dB
reference is +13
dBm .
Front-panel
level
range :
Continuously adjustable
from +13
dBm
to
-97
dBm
in
10
dB steps
and
an 11
dB
VERNIER .
Reverse power protection is also
provided by this option .
Option 1B Program Level Range :
0 to 90 dB
in
10 dB steps (programmed via rear-panel
plug) .
0 dB reference set by
front-panel
attenuators .
Remote control of CW/AM mode
is also provided .
Front-panel Level Range :
See Section 1 .2 .2 .
Reverse
power protection is also provided
by this option.