4 .6
HARMONICS TEST
SPECIFICATION
METHOD
EQUIPMENT
Spectrum Analyzer
PROCEDURE
PERFORMANCE TESTS
12 .
Repeat steps 3 through 11 using the standard attenuator
pads and the Signal Generator OUTPUT step attenuator settings
indicated in the following table .
NOTE :
To test the OUTPUT step
attenuator below -60 dBm an
RF amplifier
(>20 dB gain) is required .
Insert the 26 dB
wideband amplifier
between the standard attenuator pad and
the spectrum analyzer
(Figure 4-1) .
The allowable
error
for the -90 dBm
setting (step 11)
is ±1 .5 dB .
The OUTPUT
step attenuator can be tested down to the -130 dBm position
if a 40 dB RF amplifier is used and if precautions are taken
to properly shield the RF output from the Signal Generator .
Harmonics Outputs
>30 dB below fundamental from 10 to 520 MHz
>26 dB below fundamental from 1 to 10 MHz
A spectrum
analyzer is
used to measure harmonics
in
the
frequency range of the Signal Generator
at +13 and +3
dBm
output levels .
HP8554L/8552B/141T
1 .
Set the Signal Generator controls as follows
FREQUENCY selector
001 .000 MHz
MODULATION MODE
CW
MODULATION FREQ
(Inactive)
MODULATION FM/AM
(Inactive)
OUTPUT VERNIER
Fully Clockwise
OUTPUT step attenuator
+10 dBm
2 .
Connect the
Signal Generator
RF out connector
to the
RF input of the spectrum analyzer .
3 .
Set the spectrum analyzer
to measure the harmonic dis-
tortion of the Signal Generator for fundamental frequencies
between 1
and 10 MHz .
Set the bandwidth to 100 kHz,
the
4- 11
Steps 4 and 8
Attenuator pad
dB
-- -
Step 7
OUTPUT Step Attenuator
dBm setting
Step 11
Record Error on
Line of PTR
10
-10
12
20
-20
13
30
-30
14
60
-60
15
90
-90
16