Wavetek 3000 Instruction Manual page 47

Signal generator
Table of Contents

Advertisement

4 .5 .1
Output Meter Accuracy Test
EQUIPMENT
Power Meter
and Sensor
PROCEDURE
Note that no step attenuator pads are active in the +10 dBm
and 0 dBm positions .
A leveled pin-diode attenuator reduces
the output
level by 10 dB
in all
positions of the output
step attenuator
below +10 dBm .
The output level over the
entire range of +13 dBm to -137 dBm including a 10 dB vernier
is controlled by the pin leveler system .
The output step attenuator error is measured by an
RF sub-
stitution method .
Each of the five pads in the output step
attenuator is
measured at
520 MHz .
The second 30 dB pad
and the 40
dB pad are measured in
combination with
other
pads .
A reference output level is set with a power meter .
A reference trace is obtained with a spectrum analyzer and a
standard attenuator pad .
The standard pad is removed
and
the output step
attenuator position to be measured is sub-
stituted .
The spectrum analyzer trace is
returned to the
reference level
by resetting
the Signal Generator
output
level .
The
resulting Signal
Generator
output level
is
measured and compared to the original power meter reference
level .
A 26 dB RF
amplifier is required to
boost signal
levels below the -60 dBm level .
HP435A/8481A
1 .
Set the Signal Generator controls
FREQUENCY selector
MODULATION MODE
MODULATION FREQ
MODULATION FM/AM
OUTPUT VERNIER
OUTPUT step attenuator
power
cates
NOTE :
PERFORMANCE TESTS
as follows :
050 .000 MHz
CW
(Inactive in CW MODE)
(Inactive in CW MODE)
Fully Clockwise
+10 dBm
Set the
sensor
2 .
Calibrate the
power meter
and power sensor .
power meter to the +15 dBm range .
Connect the power
to the Signal Generator RF out connector .
(When reading the
meter,
set the range switch
so that the meter indi-
between 0 and -5 dBm) .
The Signal Generator indicated output level is equal
to the sum of the OUTPUT meter reading and the step attenuator
setting .
The difference
between
the actual power
meter
reading and the indicated output
level is the OUTPUT meter
error .
For example, the indicated output level is +13 dBm
for an
OUTPUT meter reading
of +3 dBm and
an OUTPUT step
attenuator setting of +10 dBm .
If the power meter reading
is +13 .15 dBm, the OUTPUT meter error is +0 .15 dB .
4-7