PERFORMANCE TESTS
METHOD
4 .14
FM DEVIATION ACCURACY TEST
SPECIFICATION
the meter response to fast, the function switch to +AM, the
meter range switch to 100 and the filter bandwidth to 50 Hz
to 15 kHz .
4 .
Adjust
the Signal Generator
MODULATION FM/AM
control
for a modulation meter reading of 70% AM .
Set the modulation
meter function switch
to -AM, and
observe the
modulation
meter reading .
Readjust the MODULATION FM/AM control until
the average of the two modulation meter readings in +AM and
-AM positions
of the modulation
meter function switch
is
equal to 70% AM .
5 .
Calibrate the distortion
analyzer and measure the dis-
tortion .
The distortion should be less than 3% .
Record the
distortion on line 34 of the PTR .
6 .
Adjust the Signal Generator MODULATION FM/AM control as
in step 4 until the average of the modulation meter readings
in +AM and -AM positions of the modulation function
switch
is equal to 90% AM .
7 .
Calibrate the distortion
analyzer and measure the dis-
tortion .
The distortion should
be less than 5% .
Record
the distortion on line 35 of the PTR .
Deviation Accuracy
±250 Hz on FMx1 range
±35 kHz on FMx100 range
The deviation is measured in both FM modes using an internal
DC voltage
equal to the peak
of the internal
sine
wave
voltages .
A frequency counter is used to measure the maxi-
mum deviation in both FM modes .
EQUIPMENT
Frequency Counter
HP5300B/5303B
PROCEDURE
1 .
Set the Signal Generator controls as follows :
FREQUENCY selector
050 .000 MHz
MODULATION MODE
FMx1
MODULATION FREQ
VERNIER
MODULATION FM/AM
5 kHz on FM scale
OUTPUT VERNIER
Fully Clockwise
OUTPUT step attenuator
+10 dBm
2 .
Connect
the 50 ohm
input of the
frequency counter to
the Signal Generator RF out connector .