Signal generator
Table of Contents

Advertisement

2 .
Connect the equipment as shown in Figure 4-10 .
Model 3000
Signal Generator
26 dB RF Amplifier
Spectrum Analyzer
RF Out
Input
00
Step 4
Output
Figure 4-10 .
RF Leakage Setup
3 .
Set the
spectrum analyzer
bandwidth
to 100 kHz,
the
scan width
to 0 .5
MHz/div,
the video
filter to
100 Hz,
to 0 dB and the
log reference level
Center the sig-
control .
Cali-
at the -31 dBm
the input attenuation
to -50 dBm with a 10 dB/div vertical scale .
nal in the display using the center frequency
brate the analyzer
for the -107 dBm signal
graticule using the log reference controls .
PFRFORMANCF TESTS
RF Input
4 .
Disconnect the
RF amplifier from the Signal Generator,
and connect a 100 dB attenuator pad (10+20+30+40)
to
RF out connector of
the Signal
Generator .
Terminate
attenuator in a 50 ohm load .
the
the
5 .
Set the
Signal Generator OUTPUT step attenuator to -10
dBm, and the OUTPUT VERNIER to a +3 dBm reading on the OUTPUT
meter .
6 .
Connect the loop .probe to the input of the RF amplifier .
Move the loop probe over the surfaces of the Signal Genera-
tor
with
the two-turn
loop at
a one-inch distance .
The
signal plus noise should be less than the -107 dBm reference
(step 2) .
Record the maximum reading in dBm on line 46 of
the PTR .