Replace the table shown with the following table :
MANUAL CHANGES - MODEL 3000 (Cont'd)
Section 4 .5, OUTPUT LEVEL ACCURACY TESTS :
Change
the
third
paragraph
of
the
METHOD
flatness is measured in 10 MHz steps between
+3, and -7 dBm output levels ."
Section 4 .5 .2, FLATNESS TEST :
Make the following changes in Steps 1, 4, 5,
section :
Step 1 .
Delete the line :
"FREQUENCY selector
050 .000 MHz"
Step 7 . Change the first sentence to read :
"Set the OUTPUT step attenuator to 0 dBm ."
section
to
read
"The
10 and 520 MHz at +12,
Step 5 .
Change to read :
"Set the OUTPUT VERNIER for a +3 dBm power meter reading ."
Change the last paragraph in the METHOD section as follows :
The third sentence in the paragraph must read "The two 35 dB pads
are measured in combination with other pads ."
and 7 of the PROCEDURE
Step 4 .
Replace the paragraph with the following :
Set the Signal Generator Frequency selector to read 10 MHz and step
in 10 MHz steps between 10 and 520 MHz . Observe the change in power
meter readings . The total deviation shall be not greater than 1 .5 dB
overall .
Record
the
maximum
change
to
the
nearest
.05
dB
(1/4
division) on line 9 of the PTR .
OUTPUT STEP
ATTENUATOR POSITION
dBm
+10
I
1 10
I
(ACTIVE
dB
STEP
20
I
dB
ATTENUATOR
1 30
I
dB
PADS
1 35
I
(%)
dB
1 35
I
dB
0
I
I
I
-
10
I
x
I
I
-
20
I
I
x
I
-30
I
I
I
x
-40
I
x
I
I
x
-50
I
I
x
I
x
-
60
I
x
I
x
I
x
-
70
I
I
I
I
x
I
x
- 80
I
x
I
I
I
x
I
x
-
90
I
I
x
I
I
x
I
x
-100
I
I
I
x
I
x
I
x
-110
I
x
I
I
x
I
x
I
x
-120
I
I
x
I
x
I
x
I
x
-130
I
x
I
x
I
x
I
x
I
x

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents