Wavetek 3000 Instruction Manual page 58

Signal generator
Table of Contents

Advertisement

Figure 4-3 .
Percent AM Accuracy Setup
3 .
Adjust
the Signal Generator
MODULATION FM/AM
control
to its maximum up position .
4 .
The digital
multimeter should
read 5 .000 ±0 .020 volts
DC .
If the voltage is within limits,
continue to step 5 .
If out of limits, the voltage should be recalibrated .
5 .
Adjust the Signal Generator MODULATION FM/AM control to
30% AM.
6 .
The digital
multimeter should
read between 1 .300
and
1 .700 volts DC .
Record the reading on .line 20 of the PTR.
7 .
Set the
Signal Generator
MODULATION FM/AM
control to
90% AM .
8 .
The digital
multimeter should read between
4 .300
and
4 .700 volts DC .
Record the reading on line 30 of the PTR .
9 .
Adjust the Signal Generator MODULATION FM/AM control to
OI AM .
NOTE :
This concludes the MODULATION FM/AM control accuracy
test .
As many other points may be tested as desired .
10 .
Set the modulation meter to read %AM at 520 MHz .
Set
the meter range switch to 100,
the RF input attenuation to
10 dB,
the IF bandwidth to ±400 Hz,
the meter response to
fast,
the function switch to +AM
and the filter bandwidth
to 50 Hz-15 kHz .
11 .
Adjust the
Signal Generator MODULATION FM/AM
control
for a reading of 1 .500 ±0 .003 volts DC on the digital multi-
meter .
Set the MODULATION FREQ switch to 1 kHz and the MOD-
ULATION MODE switch to AM .