OPERATION
2 .6 .3
RF Output Check
Connect the equipment as shown in Figure
2-4 .
The 10 MHz signal must beat least
0 .9 V p-p (a high frequency oscilloscope
must be used for these checks) .
2 .6 .4
AM Modulation Check (1000 Hz)
Switch the mode switch to AM .
Move the
AM/FM vernier
up to the 50% modulation
point .
Verify that the AM envelope dis-
played on the oscilloscope showsa peak-
to-valley
voltage difference
of about
.45 V and a period of 1 ms .
(See Figure
2 .6 .6
FMxl Check
Switch
the mode switch to FMxl .
Move
the AM/FM Vernier up and down .
Verify
that the oscilloscope
shows an FM dis-
play (See Figure 2-6) .
2-6
Figure 2-5 .
AM Modulation
Figure 2-6 .
FM Modulation
2 .6 .7
FMx100 Check
Switch
the mode
switch to FMx100
and
repeat the above check .
2 .6 .8
Frequency Vernier Check
Leaving the
mode switch in
the FMx100
position place the frequency switch
in
the Vernier position .
Verify that moving
the AM/FM Vernier from 0 to 5 kHz shows
an increase in frequency on the oscillo-
scope .
2 .6 .10
Attenuation Check
Switch
the mode switch to CW .
Verify
that the output vernier
and attenuator
controls change
the
amplitude of
the
2-5) .
2 .6 .5
AM Modulation Check (400 Hz)
Move the frequency switch to the 400 Hz
2 .6 .9
Vernier Output Check
Switch the
mode switch to
tion .
Verify
that
moving
the AM posi-
the AM/FM
position .
Verify the AM envelope per-
Vernier from 0 to 50 shows
an increase
iod is 2 .5 ms .
in output amplitude .
(NOTE :
The unlevel
light
may come
on during
this test .)