Frequency Range And Resolution Test - Wavetek 3000 Instruction Manual

Signal generator
Table of Contents

Advertisement

PERFORMANCE TESTS
Power Meter
Range :
10 to >520 MHz
HP435A/8481A
Input Level :
-7 to +13 dBm
Accuracy :
t1I FS
Modulation Meter
Range :
5 to >520 MHz
AFM2 Radiometer
Residual FM :
<100 Hz (rms)
(quiet room)
Residual AM :
<0 .1% (rms)
(in CW)
AM Accuracy :
±(2%R+1IFS)
Oscilloscope
Range :
DC to 2 MHz
Tektronix D10/
Sensitivity :
2 V/cm (AC coupled)
5A18N/5B10N
Spectrum Analyzer
Range :
500 kHz to 1200 MHz
HP8554L/8552B/
Display :
2 dB log and 10 dB log
141T
4 .2
FREQUENCY RANGE AND RESOLUTION TEST
SPECIFICATION
4-2
Range
1 MHz to 520 MHz selectable in 1 kHz steps .
Resolution
1 kHz
TABLE 4-1 .
(Cont'd)
METHOD
A frequency counter is used
to measure the frequency range
and the frequency resolution .
All frequencies in CW and AM
modes between 1 and 520 MHz are selected by front-panel lever/
indicator switches .
Each of the digits of the frequency se-
lector
(a total of 56) will be tested .
The 0 through 9 kHz
digits provide 1 kHz resolution .
Precision Attenuator
Pads
10, 20, 30 and 40 dB
Weinschel 50-10,
50-20, 50-30,
and 50-40
Wideband Amplifier
Range ;
1 to 520 MHz
HP8447D
Gain :
26 dB
Impedance : 50 ohm
Sweep/Signal Generator
Range :
1 to 520 MHz
Wavetek 2001
VSWR Bridge
5 to 525 MHz, 50 ohm
Wiltron 60N50
50 dB directivity
Coaxial Short
Type N female
HP11511A
50 ohm Load
BNC
HP11593A
Loop Probe
See Figure 4-9 .

Hide quick links:

Advertisement

Table of Contents
loading