Performance Tests - Wavetek 3000 Instruction Manual

Signal generator
Table of Contents

Advertisement

4 .1
INTRODUCTION
The purpose of the performance tests
in
the following
paragraphs is
to verify
that the Model
3000 Signal
Generator
meets its published specifications (par-
agraph 1 .2) .
Individual
performance
tests consist of :
the specification to
be verified,
the method of
testing, a
list of equipment required and a detailed
test procedure including in
some cases
a simplified setup drawing .
Critical
specifications for
each item
of test
equipment are listed
in Table
4-1 of Recommended Test Equipment .
Ex-
cept as detailed settings of test equip-
ment apply to performance test procedures,
all other test
equipment operating de-
tails are omitted .
The Signal
Generator
should have
its
top and bottom covers installed for the
performance
tests .
All of
the
tests
TABLE 4-1 .
RECOMMENDED TEST EQUIPMENT
FOR MODEL 3000 PERFORMANCE TESTS
SECTION
4
PERFORMANCE TESTS
can be performed
without access to the
internal
controls .
Before
applying
power to the Signal Generator see
Sec-
tion 2 for details of electrical instal-
lation .
The
line
voltage
should be
maintained at 115 or 230 volts ±10%, 50
or 60 Hz
throughout
the
tests .
The
performance test procedures
are
begun
after a two-hour
minimum warmup of the
Signal Generator
in a
+20 to
+30 0
C
ambient temperature range .
A copy of the
Performance Test
record
(PTR)
is provided
at the end
of this
section
for convenience
in
recording
the performance of the Model 3000 during
performance
tests .
It can be
filled
out and
used as a permanent record for
incoming
inspection or it can be
used
as a guide for routine performance test
ing .
The PTR lists the paragraph, test,
basic control settings and limits .
All
of the tests refer to this test record .
INSTRUMENT
Digital Multimeter
10 VDC :
Distortion Analyzer
Range :
CRITICAL REQUIREMENT
Frequency Counter
Range :
to 525 MHz
Function Generator
Level :
10 Vp-p
600 ohm load
Range :
>0 .2 Hz
Distortion : <1%
t(0 .07IR+0 .02%FS)
5 Hz to >25 kHz
RECOMMENDED
Keithley 179
HP334A
HP5300B/5303B
sine wave into
Wavetek 130
to >25 kHz