Power off behaviour

After the system is stopped or if the emergency button is activated, this is the
behaviour of the system:
Moving the ceiling suspension when the power is off, can cause damage to the
equipment.
WARNING:
The brakes for longitudinal and transversal movement of the
tabletop are released. The tabletop can move freely in
longitudinal and transversal direction when minimal force is
applied. If the patient is still on the radiographic table, he may
need help to step down from the table.
WARNING:
The brakes of the ceiling suspension movement are activated. If
the ceiling suspension makes it impossible for the patient to step
down from the table, the ceiling suspension can be moved when
sufficient force is applied.