Generator Technical Data (Spellman)

Manufacturer Spellman High Voltage Electronics GmbH
Supported
Models
Power rating
for 400 VAC
Power rating
for 480 VAC
Max. Power
Power Output
(at 0.1s)
kV-Range
mAs-Range
mAs-Range
(tomosynthe-
sis exposures)
mA-Range
ms-Range
Power line
400 V Y-
source
Power line
400/480 V
Delta-source
Dimensions
Weight
Duty cycle
Josef-Baumann-Strasse 23
D-44805 Bochum, Germany
EDITOR HFe 501
113 A / 76 kVA
97 A / 76 kVA
50 kW
625mA: 80kVp
500mA: 100kVp
400mA: 125kVp
330mA: 150kVp
40-150 kV
0.5-600 mAs
0.1-20 mAs
Maximum dose (air kerma) is reached at SID 100 cm, 150
kV, 16 mAs
10-650 mA
1-6300 ms
400V 3N~ PE (Y) 50/60 Hz
400/480V 3~PE (delta without N) 50/60Hz
The power setting is selected during installation and printed
on the type label.
97 cm x 55 cm x 63 cm (HxWxD)
129 kg
The Generator duty cycle is continuous, but limits should be
set during installation depending on the capacity of the X-
ray tube.
DR 600 | Product Information | 283
EDITOR HFe 601
144 A / 96 kVA
124 A / 96 kVA
65 kW
800mA: 80kVp
650mA: 100kVp
520mA: 125kVp
430mA: 150kVp
10-800 mA
EDITOR HFe 801
180A / 120 kVA
154 A / 120 kVA
80 kW
800mA: 80kVp
800mA: 100kVp
640mA: 125kVp
530mA: 150kVp
3261K EN 20210708 0808