DR detector and X-ray tube head alignment

The tube head display can show a readout of the tilting angle of the DR
detector and of the X-ray tube head.
When both angles differ by less than one degree, they're displayed in green.
The exposure is perpendicular if both angles are displayed in green and the
tilting axes of the DR detector and of the X-ray tube head are parallell. The
alignment of the tilting axes must be visually verified by the user.
The relative angle between the DR detector and the horizontal surface.
1.
The rotation of the DR detector around the vertical axis. This angle is not
2.
measured and must be visually verified by the user.
If the angle is very small (smaller than 3 degrees), it is hard to visually verify
the alignment of the tilting axes. Use the light field of the collimator to verify
the alignment of the X-ray tube head to the DR detector.
Depending on the orientation, the angle may be displayed as a negative
number. The number sign is ignored for comparing the angles.
The availability of the alignment feature depends on the DR detector model
and on the product license.
DR 600 | Software Console and Tube Head Display | 155
The relative angle between the DR detector and the horizon-
tal surface.
The relative angle between the X-ray tube and the vertical ax-
is.
2
CAUTION:
The alignment readout may not be available, e.g. if the position
of the detector is not stable due to movement of the patient. A
skilled operator can do the alignment visually.
1
3261K EN 20210708 0808