DR 600 | Ceiling Suspension | 191

Semi-automatic collimation mode

The semi-automatic collimation mode is activated if any of following
conditions applies:
• the tube head unit is rotated out of the center position
• the SID on the radiographic table is not within 90 cm to 130 cm
• the SID on the radiographic wall stand is not within 90 cm to 205 cm
• the tube head unit is not centered to bucky
In semi-automatic collimation mode the registration of the cassette or detector
format in the bucky is stopped, but the collimation is still adapted when the
SID is changed. The user can adjust the collimation manually.
Figure 83: Indication on the tube head display for semi-automatic collimation
mode