Automatic centering

The automatic centering functionality is intended to center the X-ray tube
head unit over the detector or cassette in the bucky on radiographic table or
radiographic wall stand.
On the radiographic table automatic centering is limited to longitudinal and
transversal movements.
On the radiographic wall stand automatic centering is limited to transversal
and vertical movement relative to the radiographic wall stand, so SID or
angulation of the tube head unit is not changed by this function.
Automatic centering is used to make sure the X-ray tube head and the bucky
are aligned after applying manual movements.
Automatically centering the X-ray tube head unit
The automatic centering button on the main screen of the tube head display
indicates if automatic centering can be performed.
Table 32: Automatic centering status
1. Press the automatic centering button on the main screen of the tube head
display.
2. Press and hold the automatic positioning button.
The status of the automatic centering is displayed on the main screen of
the tube head display:
DR 600 | Software Console and Tube Head Display | 171
Automatic centering can be performed, but is not activated.
Automatic centering cannot be performed.
Make sure that the X-ray tube is pointing to the detector and
that it is within a range of 50 cm from its center position.
3261K EN 20210708 0808