Cassette and detector formats

To adjust the side clamps to the format of the cassette or detector, indications
are available in cm (and inch, depending on the bucky type). Corresponding
indications are printed on the wall stand cover to align the collimation area.
The large format cassette or detector (43 cm x 35 cm or 17 inch x 14 inch) can
be positioned either centered or aligned to the top of the bucky in landscape
position.
Indicators for large format cassette or detector position to the top of the
1.
bucky
Large format detector positioned to the top of the bucky
2.
Figure 103: Wall stand bucky with large format detector positioned to the top
of the bucky
1
Indicators for large format cassette or detector position to the top of the
1.
bucky
Figure 104: Indicators on bucky tray