Behavior in Case of Power Failure

Note: The description below is only applicable if an uninterruptible power
supply (UPS) is put into the CR 30-X System configuration. For more
information about the CR 30-X System, refer to the CR 30-X System User
Manual, document 2385.
In case of a power failure, the system is still connected to the UPS. Two
situations are possible:
Power failure after cassette insertion and before identification with the NX
workstation. CR 30-X pushes the image plate back in the cassette without
scanning and releases the cassette. After the power supply returns, the
cassette must be inserted into the digitizer and identified again to read out
the image.
Power failure after identification with the NX workstation. The image plate
is scanned and erased as usual. The scan cycle finishes when the cassette is
released. If the power supply is still not available, the digitizer will refuse
scanning other cassettes.