Diagnostics On The Cometer; Hart Version Diagnostics On The Hht; Hart Version Diagnostics On Thepc - ABB 2600T Pressure Transmitter Series User Manual

Table of Contents

Advertisement

364xS User's Manual
Corrective actions: Check the ambient temperature. If it is within the limits see 1, 2, 3.
Consistency Check
The secondary electronic and the primary electronics have not been working together before.
Corrective actions: If this message appears during a replacement, please follow the replacement instruction
in the relevant section of this manual. If the error persists see 1, 2, 3 or 4
Config. Error
Bad configuration of range and/or transfer function.
Corrective actions: Please follow the setting configuration instruction in the relevant section of the this
manual. If the error persists see 1, 2, 3 or 4
Sensor Elec. Fail
There are internal inconsistencies in the memory of the primary electronic. This test is performed at the start
up.
Corrective actions: 1, 2, 4
Tot. Config Error
The configuration of the totalizer is not correct
Corrective actions: Please follow the totalizer instruction in the relevant section of the this manual. If the
error persists see 1, 2, 3 or 4
Common corrective actions:
1.
Switch the transmitter off and on to verify if the message appears again
2.
With the HART command 48 it is possible to obtain additional information on the failed component
3.
If the message persists the secondary electronic is suspected to have a failure. If the output is in a
failure status (high or low) the fault is severe and it is necessary to substitute the secondary electronic.
Otherwise the failure is minor, the signal is still valid but with a lower accuracy. The secondary
electronic can be substituted when convenient.
4.
If the message persists, the primary electronic could be damaged. If the output is in a failure status
(high or low) the fault is severe and it is necessary to substitute the transducer. Otherwise the failure is
minor, the signal is still valid but with a lower accuracy. The transducer can be substituted when
convenient.
5.
Check the process and ambient conditions (e.g.: pressure, process temperature, ambient temperature
static pressure, etc.) against the sensor and electronics limits. If the conditions are close or outside the
limits, these improper instrument utilization can have seriously compromised the instrument
functionality. If the conditions are well inside the limits see the corrective actions of point 4 or 3.

Diagnostics on the CoMeter

The diagnostics messages on the Cometer and Prometer when installed on the 695 Field Indicator are
available. Please refer to the device manual for the right operations.

Hart version diagnostics on the HHT

Hart version diagnostics on the Hand Held Terminal are available. Please refer to the device manual for the
right operations.

Hart version diagnostics on thePC

Hart version diagnostics on the PC are available using dedicated Software. ABB can provide the references
and the Software. Please refer to the Software instructions for the right operations.