Other Fault And Alarm Messages; Maintenance - M&C CSS-VC1 Instruction Manual

19", wall mount or portable in carrying case with software version
Hide thumbs Also See for CSS-VC1:
Table of Contents

Advertisement

This prevents liquid pumped along to the analyzer and thus damaging! The triggered alarm (red LED) must be
acknowledged by pressing any key. If the error is not corrected, the error screen comes back after 60 s.
18.4

OTHER FAULT AND ALARM MESSAGES

Error message
Pump-Warn.: 1, 2
Maintenance needed!
Temp.2
19

MAINTENANCE

Before performing maintenance work, the system and process-specific safety measures must be observed!
Warning
The maintenance intervals depend on the process conditions and therefore need to be determined for specific
applications.
All serviceable parts are mounted in an easily accessible location in the front of the compact gas conditioning unit
CSS-VC...
Change filter elements of the dust and aerosol filters (FPF-0,1GF/FPF+/universal filter/aerosol filter) when there
is too little flow or after visual inspection (see chapter 19.1);
www.mc-techgroup.com
Alarm
Warning
x
x
-
x
x
-
Hazardous voltage. Pull the mains plug out before carrying out work on the
gas conditioning unit!
CSS-VC-1.6 | 1.02.01
LED
Error
red +
Pump 1 and/or 2 off due to an alarm (temperature,
yellow
liquid)
yellow
Maintenance interval expired
red
Pt100 for heated sample line connected