Rechargeable Battery Regeneration Function - BRAND Transferpette electronic Operating Manual

Micropipettes
Hide thumbs Also See for Transferpette electronic:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 45
10 Maintenance
Replace rechargeable battery
a.
b.
c.
Battery indicator after reinserting a battery pack

10.4 Rechargeable battery regeneration function

Refresh function
To extend the service life and increase the performance of the batteries, the Transferpette® electronic
has a regeneration function (refresh function). This function allows the batteries to be fully discharged
and recharged under program control. To optimize the performance of the batteries, the refresh func-
tion should be used from time to time.
72
Operating manual
a.
Open the cover of the battery compartment, remove
the rechargeable battery, and pull the plug out of the
socket.
b.
Insert the plug of the new rechargeable battery into the
socket and insert the new rechargeable battery.
c.
Replace the cover of the battery compartment and
close it.
Remove the rechargeable battery from the instrument during
longer breaks in operation.
After inserting a rechargeable battery, the full capacity indi-
cator appears in the display with a flashing frame (the instru-
ment does not yet recognize the charge status at first). After
3.5 h charging time – safe complete charging of the battery –
the frame stops flashing.
Always charge for at least 3.5 h after inserting a battery! The
full charge capacity is reached after several charge/discharge
cycles!
NOTICE
Transferpette® electronic 997209

Advertisement

Table of Contents
loading

Table of Contents