5. Hygienic preparation

Note!
This product may contain disposable items. Disposable items are
intended to be used only once. So use these items only once and do
not reprocess them. Reprocessing disposable items may impair the
functionality and safety of the product and lead to unforeseeable
reactions as a result of ageing, embrittlement, wear, thermal load, the
effects of chemical processes, etc.
5.1 Critical times
• Particle filter 7 must be changed after
24 h (3 nights of 8 h each).
• The housing of the bacteria filter
should be cleaned after 24 h (3 nights
of 8 h each).
• Particle filter 7 must always be
7
changed and the housing disinfected
or sterilized whenever there is a new
patient.
• The pressure measuring tube of the
tube system should be changed every
four weeks on hygiene grounds.
Hygienic preparation
25
EN