5. Hygienic preparation

Valve cage with
expiration cone
Patient valve
Safety valve
(relief valve)
Squeeze
bag
Intake port inner
part with
diaphragm
Intake port cov-
er with holding
ring
COMBIBAG should undergo hygienic purification
after every use.
Upon completion of the purification process, carry
out a functional test.
Individual parts of the COMBIBAG resuscitator are
made of silicon or polysufone.
Use soapy water to
der running water. For cleaning purposes, take apart
the bag, separating the pieces into components as
described.
Warning:
When assembling or disassembling
COMBIBAG, do not use any sharp-edged objects.
The patient valve does not have to be dismantled in
order to be cleaned!
Should you choose to dismantle the valve anyway,
do not unscrew the cap of the valve
change the pressure setting of the valve.
disinfect
To
the equipment, use a disinfectant that is
suitable for rubber and polysulfone. Be sure to ob-
serve the instructions for the disinfectant. We rec-
ommend GIGASEPT FF.
Sterilization with steam, hot air or gas should be car-
ried out according to standard practices after careful
cleaning of the individual parts. All parts can be au-
toclaved at 134 °C.
The oxygen reservoir
be cleaned and disinfected.
clean
COMBIBAG and rinse un-
since that could
cannot be sterilized