5. Hygienic preparation

5.1 Intervals

This product may contain disposable items. Disposable items are intended to be used
only once. So use these items only once and do not reprocess them. Reprocessing dispos-
able items may impair the functionality and safety of the product and lead to unforeseeable
reactions as a result of ageing, embrittlement, wear, thermal load, the effects of chemical
processes, etc.
Clean the device as described in Section "5.2 Cleaning" as required, e.g. after use in a
dusty environment.
Disinfect the device as described in Section "5.3 Disinfecting" as required, in particular
when changing patients or in the event of direct contact between the patient and the
device.

5.2 Cleaning

Warning!
Risk of fire. Do not use cleaning agents. Cleaning agents containing alcohol or
moisturizers form a flammable mixture in combination with oxygen.
Caution!
Device damage possible. Ensure that no liquids get into the device, otherwise
there may be damage to the device.
• If external cleaning is required, use only a clean cloth. The cloth can be
dry or moistened with clean water.

5.3 Disinfecting

Warning!
Risk of injury. Ensure that no disinfectant gets into the tubes, otherwise the disin-
fectant will get into the patient's respiratory tract together with the oxygen.
Device damage possible. Ensure that no liquids get into the device, otherwise
there may be damage to the device.
Hygienic preparation
EN
19