2. For nominal 230VAC operation, proceed as follows:
K
Change the line fuse to a 80mA slow blow type
fuse;
K
Remove the two 120V(marked as 110V) soldered
links on the module (see fig. b);
K
Solder one link provided for the 230V operation
(fig. a);
K
Remove the existing plug from the power cord
and strip the wires to attach the appropriate CEE7
power plug.
Power cord with CEE 7 plug
The wires of the power cord are colored in
accordance with the following code :
Green-and-yellow
Blue
Brown
3. For nominal 120VAC operation, proceed as follows:
K
Change the line fuse to a 160mA slow blow type
fuse;
K
Remove the 230V soldered link on the module
(see fig. a);
K
Solder two links (marked as 110V) provided for
120V operation (see fig. b);
K
Remove the existing plug from the power cord
and strip the wires to attach the appropriate power
plug, ANSI 73.11.
Power cord with ANSI 73.11 plug
The wires of the power cord are colored in accord-
ance with the following code :
Green-and-yellow
White
Black
5975636
ACTAS 2
170297
: Earth (safety earth)
: Neutral
: Line (Live)
: Earth (safety earth)
: Neutral
: Line (Live)
230VAC operation
Fig. a
line fuse
T80mA
230 V link
110 V link
CEE 7 plug
120VAC operation
Fig. b
line fuse
T160mA
T80mA
230 V link
110 V link
ANSI 73.11 plug
ACTAS set-up
110 V link
110 V link