6.7.2 Manual span calibration

SPAN
ENT
plied.
Note
For the Ch(components) in thich "both" is set
in the calibration range setting (Refer to Section
6.2.3) - span calibration is completed for both
ranges.
ENT
Note: For the Ch (component) for
which"AR" is selected in Section
6.1.1 "Setting range switch mode,"
the cursor automatically moves
to the range selected in "Setting
of auto calibration component/
range" (Section 6.2.4), and cali-
bration is carried out within that
range.
To close "Span Calibration"
To close the "Span Calibration" or cancel
this mode midway, press the
A previous screen will return.
INZ-TN2ZPAH-E
SPAN Cal.
-
SPAN Cal.
key and press
-
SPAN Cal.
ESC
key.
51
SPAN
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-500.0
Ch2
CO
Range 2 0-2000
(
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-500.0
Ch2
Range 2 0-2000
CO
ENT
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-500.0
Ch2
Range 2 0-2000
CO
ENT
To Measurement screen after
executing Manual Span Calibration
mea-
4
vol%
0.00
vol%
ppm
0.0
ppm
)
ENT
vol%
0.00
vol%
ppm
0.0
ppm
vol%
0.0
vol%
ppm
490.0
ppm