Site Preparation; Cooling Considerations - Agilent Technologies HP 5890 II Series Manual

Gas chromatograph
Table of Contents

Advertisement

Installation

Site preparation

Site preparation
If not already done, refer to Site Preparation in this manual for
information concerning proper site preparation.

Cooling considerations

Cooling: The instrument is cooled convectively: ambient air enters
through vents in side panels and from beneath the instrument. Warmed
air exits through slots in top, rear, and side panels. Since ventilation slots
in instrument top, side, and rear cover panels provide instrument cooling,
materials that might obstruct air flow must not be placed beneath or on
top of the instrument.
For proper cooling and general safety, always operate the instrument
Caution
with cover panels properly installed.
Oven Exhaust: Heated air (up to 450
through a vent in the rear. Allow at least 25 cm (10 inches) clearance
behind the instrument to dissipate the heated air. Also, place nothing
behind the instrument that might be damaged by the heated air exhaust.
Temperature•sensitiveitems (gas cylinders, chemicals, regulators, etc.)
WARNING
should not be located in the path of the heated exhaust. Also, exercise
care in working at the rear of the instrument during cool•downcycles
to prevent possible burns.
^
C) from the column oven exits
21