Changing The Flow Restrictor - Agilent Technologies HP 5890 II Series Manual

Gas chromatograph
Table of Contents

Advertisement

Installation
Connecting supply gases

Changing the flow restrictor

You need the following tools to complete this task: a 7/16•inchopen end
wrench, a short flathead screwdriver, a pair of tweezers, and the flow
restrictor that fits the flow range you need for your analysis.
If the gas lines are already connected to the instrument, turn off the
carrier gas source and remove the gas line from the carrier gas inlet
fitting with a 7/16•inchopen end wrench.
1. Select the flow restrictor you need for your analysis from the table.
Part number
19362-60515
19362-60525
19362-60535
19362-60545
19362-60555
*Note: Based on Helium gas
2. Remove the screw that holds down the plate on the flow controller.
26
These restrictors are used in packed and purged•packedinlet systems
(standard flow controller part number 19362•60565).There is a spare
restrictor in the ship kit (part number 19362•60535)with a flow range
of 0 to 110 ml/min.
Flow range
(ml/min)
0 to 20
0 to 60
0 to 110
0 to 400
0 to 750
Both the flow restrictor and brass carrier gas inlet fitting are held
onto the block by this plate. Set the screw, plate, carrier gas inlet
fitting aside. You will replace them in step 5.
Supply pressure*
50
50
50
50
50
Colored dot code
red & purple
blue & purple
green & purple
black & purple
yellow & purple