Agilent Technologies HP 5890 II Series Manual page 30

Gas chromatograph
Table of Contents

Advertisement

Installation
Connecting supply gases
Assemble fitting,
bottom tubing
C
C
C
C
For a carrier gas, its tube is connected at the fitting on the
appropriate mass flow controller labeled IN.
For a capillary makeup gas, its tube is connected at the fitting on
the appropriate flow manifold block labeled FID—AUX for an FID or
NPD, or TCD—AUX for a TCD or ECD.
1
For H
and air for an FID or NPD , their tubes are connected at
fittings on the appropriate flow manifold block labeled, respectively,
FID—HYD and FID—AIR.
For TCD reference gas, its tube is connected at the fitting on the
appropriate flow manifold block labeled TCD—REF.
Once all connections are made, proceed to Setting initial supply
pressures and leak testing.
Hand-tighten, then
wrench-tighten
29