Carrier Gas Miser; Making Pipe Thread Connections - Agilent Technologies HP 5890 II Series Manual

Gas chromatograph
Table of Contents

Advertisement

Site Preparation
Assembling the gas plumbing

Carrier gas miser

A carrier gas miser reduces the use of carrier gas during off•hours.It
reduces two channels of carrier gas flow to a low purging rate during
off•hours,without disturbing the pressure regulator or flow controller
settings.
Carrier Gas Miser

Making pipe thread connections

Avoid pipe thread connections if possible. If they must be used, seal them
with Teflon tape. Do not use pipe dope to seal the threads; these
mixtures contain volatile materials that will contaminate the tubing.
Teflon tape is available in two grades. The less expensive hardware
store grade contains a small amount of volatile material. In most cases
this does no harm, but it will interfere strongly (noise, elevated baseline,
reduced sensitivity and linearity) with electron capture detectors. An
instrument grade (the name may vary) has had these volatiles removed.
The better grade tape is recommended for all chromatographic fittings.
16