Installation
Connecting signal output cables
Each device to be included in INET communications is connected in
series with the next, forming a loop; thus, INET OUT on a given device
must be connected to INET IN on the next. The loop must be continuous,
and all devices must be on for the loop to function.
Typical Inet Loop System
To install the INET communications cable:
1. Locate INSTRUMENT NETWORK IN (male) and OUT (female)
2. Locate INET IN/OUT receptacles on other devices to be included in the
3. Install an INET cable between the OUT receptacle on the HP 5890
42
Sampler
5890
SERIES
II
HP 5890 SERIES II Gas Chromatograph
receptacles at the top of the HP 5890 (see the Locations for Signal
Cable Connections figure in this section).
loop.
and the IN receptacle on the next device included in the loop.
INET Cable
OUT
IN
IN
OUT
Controller
OUT
IN
Controller & Integrator
INET Cabl