Procedure
u Close the lens aperture.
v Set [Pixel Black Correct] to [User].
w Execute [Pixel Black Direct], and perform calibration.
Perform PRNU calibration.
4
The camera is equipped with a PRNU correction function. The function allows you to reduce PRNU generated
by the sensor. To use the function, calibration must be performed within the camera and correction data must
be created beforehand.
Procedure
u Make the signal output level as flat as possible within a range of 50% to 95%.
v Set [Pixel Gain Correct] to [User].
w Execute [Pixel Gain Direct], and perform calibration.
5
Adjust the black level.
Open [Analog Control], and specify the black level in [Black Level] under [Black Level Selector].
Step 6: Configuring Various Other Settings
See "Settings List" (page 37), and configure settings as necessary.
* We recommend performing DSNU and PRNU calibration again whenever the line rate setting is changed
significantly.
Step 7: Saving the Settings
The setting values configured in Control Tool will be deleted when the camera is turned off. By storing
current setting values to user memory, you can load and recall them whenever necessary. You can
save up to three sets of user memory settings (User 1 Data to User 3 Data).
Note
The setting values are not saved to the computer (Control Tool).
User memory
Temporary memory
User 1 Data
User 2 Data
Save
User 3 Data
— 17 —
Current
setting
values
SW-8000M-PMCL