Table 3-4: IEEE & RS-232 Commands(continued)
Command
Cal Due:<tab><nn/nn/nnnn><CR>
Ground Fault Stop:<tab><Off, On, or Off - DUT Grounded<CR>
Arc Pulse Width:<tab><nnmicrosec><CR>
If the enhanced features is set to Off the format of the returned data is as above, ending
with the IEEE Address.
Note that if a feature is not available, a <CR> will be returned.
LOCKout?
View the Lockout state
CALibrate:
FULL
GCONly
DATE
STOP
Stops a measurement
MEASure
Triggers a measurement of single or sequence test.
FETCh?
Fetches the most recent measurement results. The character sequence formats are as
follows:
Normal Measurements:
<meas param name>, <test voltage>, <meas param value><units> <pass or Lo ramp, Hi
ramp, Lo fail, Hi fail, Arc fail, GC fail, Sys fail>
Note: If a measurement is stopped (by stop button or remote command)
Page 84 of 84
Function
Perform full calibration
Perform Ground Continuity only calibration
Set the date to month, day, year
the reply will be<meas param name>, <test voltage>, <meas
param value><units> STOP FAIL
If a measurement is stopped (by stop button or remote command)
the reply will be<meas param name>, <test voltage>, <meas
param value><units> STOP FAIL <error type>
where error type is one of the following:
ERROR OVERLOAD, ERROR SAFETY
SHUTDOWN or ERROR INTERLOCK OPEN
Parameter(s)
mm/dd/yyyy
Interface

Advertisement

Table of Contents
loading

This manual is also suitable for:

Guardian 2510Guardian 2520Guardian 2530