11. To set Fall Time:
Press
up/down arrow keys to increase or decrease the fall time between Off and
0.1 to 999.9sec. During fall time the programmed voltage decreases linearly to zero. No
Hi/Lo limits are checked during this time.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
Note:
Under failed conditions for high limit, low limit or arc limit the 2500 will still
execute the fall time (if programmed) during the shutdown process.
12. To set Frequency:
Press
up/down arrow keys to toggle the frequency between 50 or 60Hz
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
13. To set Gnd Continuity:
Press
up/down arrow keys to increase or decrease the resistance limit between Off and
0.01 to 10.00Ω. The ground continuity measurement is made between the G.C. and Return
connections. A ground continuity test is performed prior to a Hipot of IR test when the limit
is set to a value other than Off. Any measured value above this limit is considered a
FAIL, no low limit check is performed.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
Operation/Programming
1
Program

Fall Time:

5.0sec
1
Program
Frequency: 60Hz
1
Program
Gnd Continuity:
left arrow for previous
left arrow for previous
0.10Ω
left arrow for previous