QuadTech Guardian 2500 Series Instruction Manual page 10

Ac/dc/ir hipot tester
Table of Contents

Advertisement

Common Features
Ground Continuity:
Limits:
Ground Fault
Interrupt:
Arc Detection:
Indication:
Buzzer Level:
Time:
Standard Interfaces: RS-232
Connector: 9 pin male D-series (compatible w/Sentry Series)
Test Setups:
25 User Defined (shipped with some factory settings)
9 Multi-step (3 steps)
Connectors:
Front and Rear Connection
Front Panel Lockout: 6 Digit Password with or without setup recall
Miscellaneous:
Page 10 of 10
Specifications (continued)
Test Current:
100 mA DC +/- 10%
Range:
10mΩ to 10Ω, in 1mΩ steps
Accuracy:
+/- (5% of reading + 20mΩ)
Programmable Hi/Lo during Test Time
(Lo can be set to Off : Hipot & GC, Hi can be set to Off for IR)
Programmable Hi/Lo during Ramp Time (AC/DC Hipot only)
(Both limits can be set to Off)
Shutdown within 2msec for current imbalance > 250µA (AC)
> 400µA (DC)
Arc Level AC: 0.5mA - 15mA in 0.5mA increments
Arc Level DC: 0.5mA - 15mA in 0.5mA increments
Arc Time:
1 to 50 µsec programmable
Pass/fail display, lights, audible sound
Low, High, Off
AC Hipot:
Ramp, Test, Fall:
DC Hipot:
Ramp, Dwell, Test, Fall:
IR:
Ramp, Dwell, Test, Fall:
GC:
Test:
(Test can be set to Continuous. Ramp, Dwell, & Fall can be set to Off)
Remote:
Inputs:
Start, Stop
Characteristics: TTL active low, Pulse width >1ms
Outputs:
Pass, Fail, Under Test
Characteristics: Dry contact relay, Closed if true
36V, 100mA max
High Voltage: Locking, Amp Part # 861610-2
Return:
BNC female (Selectable for ground or virtual ground)
Continuity:
Standard Banana Socket
Continue Step on Fail
Continuous Voltage on Fail
0.1 to 999sec (+/-50ms)
0.1 to 999sec (+/-50ms)
0.1 to 999sec (+/-50ms)
0.1 to 999sec (+/-55ms)
Stop Test on Pass
Step and Increment

Advertisement

Table of Contents
loading

This manual is also suitable for:

Guardian 2510Guardian 2520Guardian 2530