Condensed Operating Instructions (continued)
Connection to S50 Using Corded Product Adapter (Part # 2000-13)
(Ground Fault Stop on the 2500 must be set for Off – DUT Grounded, and START must be
initiated from the Sentry 50)
Guardian 2510/20/30
To
9 Pin
Remote
Cable
labeled
BNC to banana
Hipot
plug adapter
Cable
labeled
Sentry 50
S50
To
9 Pin
Remote

Initiating Tests

1. Before initiating the test the 2500 should be powered up, the desired test #
recalled/programmed, measure offset zeroed, and test device connected.
2.
Display should indicate Idle mode (upper right) and the test # desired (upper
left) before initiating the test. Make sure hands are clear from the device under test and press
START (green button) to initiate the test. The HIGH VOLTAGE indicator flashing (or
constantly lit on some later units) serves as warning to the operator that high voltage is
present. The STOP button may be pressed at any time to abort a test and turn the high
voltage off.
2. The instrument will automatically begin the test cycle, showing the count down through any
programmed ramp, dwell and test times.
G.C. RETURN HV OUT
White
Banana to banana
(part of S15 Cable Set)
D-
S-
S+
D+
Black (banana)
Black (lug)
1
0.325KV
AC Tot * 0.005mA
1
0.325KV
AC Tot * 0.005mA
Ramping
6.5s
Testing
3.5s
DUT
Case
Alligator Clip, red
(part of G15 Lead Set)