Lo Limit Ramp; Dwell Time; Test Time - QuadTech Guardian 2500 Series Instruction Manual

Ac/dc/ir hipot tester
Table of Contents

Advertisement

8. To set Lo Lim Ramp (AC or DC Current only, not available with Ramp Time set
Off):
Press
up/down arrow keys to increase or decrease the limit setting between Off and
0.001mA and 14.999mA (for AC) or 0.0001mA and 7.499mA (for DC) or 0.01MΩ and
1999000MΩ (for Insulation Resistance).. This Lo Lim Ramp limit is checked only
during the programmed ramp time.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
9. To set Dwell Time (DC Current or Insulation Resistance only)
Press
up/down arrow keys to increase or decrease the dwell time between Off and
0.1 to 999.9sec (for DC Current and Insulation Resistance). No Hi/Lo Limits are checked
during the programmed dwell time.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
10. To set Test Time:
Press
up/down arrow keys to increase or decrease the test time between Continuous and
0.1 to 999.9sec. Both Hi and Lo current limits (if programmed) are checked during the
test time. When the test time is programmed for Continuous the high voltage remains
On until a failure occurs or the STOP button is pressed.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
Page 44 of 44
1
Program
Lo Lim Ramp: 10.000mA
1
Program

Dwell Time:

5.0sec
1
Program

Test Time:

5.0sec