3. To set Hi Limit:
Press
up/down arrow keys to increase or decrease the limit setting between 0.001mA
and 15.000mA (for AC Current), 0.0001mA and 7.500mA (for DC Current), 0.01MΩ and
2000000MΩ or Off
Continuity). This limit is checked only during the programmed test time. For an AC or
DC Hipot test any current above this limit is considered a FAIL and for an Insulation
Resistance measurement any value above this is considered a FAIL.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
4. To set Lo Limit:
Press
up/down arrow keys to increase or decrease the limit setting between Off and
0.001mA to 14.999mA (for AC Current), Off and 0.0001mA to 7.4999mA (for DC Current),
0.01MΩ to 1999000MΩ (for Insulation Resistance), and Off and 0.01Ω and 10.00Ω (for
Ground Continuity).
only during the programmed test time. For an AC or DC Hipot test any current below
this limit is considered a FAIL and for an Insulation Resistance measurement any value
below this is considered a FAIL.
Press
right arrow key to move to the next parameter (
parameter) or PROGRAM key to exit program mode.
Page 42 of 42
1
High Limit: 15.000mA
(for Insulation Resistance) and 0.01Ω and 10.00Ω (for Ground
The Lo limit must be set lower than the Hi Limit and is checked
1
High Limit:
12.000mA
Program
Program
Operation/Programming
left arrow for previous
left arrow for previous