Limit - QuadTech 1865 Instruction Manual

Megohmmeter/ir tester
Table of Contents

Advertisement

2.6.2.8 Limit

Setup
Figure 2-15: Limit Value (Resistance or Current)
Allows entry of a single measurement limit in scientific units (up to 4 digits with decimal
and exponent) for resistance or current depending on the results parameter selected. The
exponent must be 3 through 14 for resistance or -3 through -13 for current. This limit
determines the status of the Pass/Fail indicator and Pass/Fail display, when enabled.
When the 1865 instrument is in the resistance mode the limit is a minimum value i.e. any
value above the limit will result in a Pass, whereas a value below the limit will result in a
Fail. The opposite is true when the unit is in the current mode, the limit is a maximum
value, any value below the limit will result in a Pass and a value above the limit will
result in a Fail.
If the display parameter on the I/O Menu is selected as resistance the value entered will
be an R limit, if the display is selected as current the value entered will be an I limit. The
1865's processor will check for consistency between the parameter display selected and
limit entered.
An example of the sequence for entering 1MΩ as a limit would be:
Press [ENTER]
Press [1]
Press [e] select key
Press [6](exponent multiplier)
Press [ENTER]
Operation
I/O
Utilities
Voltage
= 1000
Charge Time
=
10
Dwell Time
= 10
Measure Time
= 10
=
Discharge Time
10
Mode
Auto Manual
Range
> >
Limit
=
1 e + 006
Stop on Pass
=
0
#
to Average
=
0
In the case where I (current) is selected rather
than R (resistance), the exponent would be
entered as a negative number.
or
e
Press [ENTER]
Press [1]
Press [0] (zero), three times
Press [e] select key
Press [3]
Press [ENTER]
Page 43 of 87

Advertisement

Table of Contents
loading

Table of Contents