Ac Voltage Verification Test - Keysight Technologies U1610/20A Service Manual

Handheld digital oscilloscope
Hide thumbs Also See for U1610/20A:
Table of Contents

Advertisement

4 On both the calibrator and the U1610/20A, set each voltage range in the order
shown in
5 Compare the measurement results to the corresponding test limits of
Table
Table 1-4
Settings for the DC voltage verification test
Calibrator setting
Function
100 mV
1 V
10 V
V DC
100 V
1000 V

AC voltage verification test

1 Put the calibrator in Standby (STBY) mode.
2 On the U1610/20A, press
to select the V AC measurement function.
3 Connect the calibrator to the U1610/20A meter terminals using a banana plug
as shown below. Ensure the polarity of the banana plug connection is correct.
U1610/20A
Keysight U1610/20A Service Guide
All manuals and user guides at all-guides.com
Table 1-4
below.
1-4.
U1610/20A setting
100 mV
1 V
10 V
100 V
1000 V
Performance Verification Tests
Minimum limit
99.85 mV
0.9986 V
9.989 V
99.89 V
998 V
>
(F1) and use the
Fluke 5520A calibrator
1
Maximum limit
100.15 mV
1.0014 V
10.011 V
100.11 V
1002 V
keys
27