Capacitance Verification Test - Keysight Technologies U1610/20A Service Manual

Handheld digital oscilloscope
Hide thumbs Also See for U1610/20A:
Table of Contents

Advertisement

1
Performance Verification Tests
Table 1-6
Settings for the resistance verification test
Calibrator setting
Function
1000 Ω
10 kΩ
100 kΩ
Resistance
1 MΩ
10 MΩ
100 MΩ

Capacitance verification test

1 Put the calibrator in Standby (STBY) mode.
2 On the U1610/20A, press
to select the capacitance measurement function.
3 Connect the calibrator to the U1610/20A meter terminals using a banana plug
as shown below. Ensure the polarity of the banana plug connection is correct.
30
All manuals and user guides at all-guides.com
U1610/20A setting
1000 Ω
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
U1610/20A
Minimum limit
996.7 Ω
9.967 kΩ
99.67 kΩ
0.994 MΩ
9.917 MΩ
98.47 MΩ
>
(F1) and use the
Fluke 5520A calibrator
Keysight U1610/20A Service Guide
Maximum limit
1003.3 Ω
10.033 kΩ
100.33 kΩ
1.006 MΩ
10.083 MΩ
101.53 MΩ
keys