Performance Verification Tests
3 Set the calibrator output to 140 V. Wait a few seconds for the measurement to
settle.
4 On the U1610/20A:
a Press
b The Y1Y2 value on the display should be within the test limits of
5 Continue to verify the voltage measurement accuracy with the remaining V/div
settings in
6 Once completed, disconnect the calibrator from the U1610/20A.
7 Repeat steps 1 to 6 for the U1610/20A channel 2.
Table 1-3
Settings for the voltage measurement accuracy verification test
U1610/20A setting
50 V/div
50 V/div
20 V/div
10 V/div
5 V/div
2 V/div
1 V/div
500 mV/div
200 mV/div
100 mV/div
50 mV/div
20 mV/div
10 mV/div
5 mV/div
2 mV/div
22
All manuals and user guides at all-guides.com
(F4) and use the
the center of the voltage trace.
below.
Table
1-3.
Calibrator setting
175 V
–175 V
140 V
70 V
35 V
14 V
7 V
3.5 V
1.4 V
700 mV
350 mV
140 mV
70 mV
35 mV
14 mV
or
key to place the Y2 cursor at
Minimum test limit
157.40 V
–192.60 V
132.96 V
66.48 V
33.24 V
13.30 V
6.65 V
3.32 V
1.33 V
664.80 mV
332.40 mV
132.96 mV
66.48 mV
33.24 mV
13.30 mV
Keysight U1610/20A Service Guide
Table 1-3
Maximum test limit
192.60 V
–157.40 V
147.04 V
73.52 V
36.76 V
14.70 V
7.35 V
3.68 V
1.47 V
735.20 mV
367.60 mV
147.04 mV
73.52 mV
36.76 mV
14.70 mV