Multimeter Performance Verification Tests; Dc Voltage Verification Test - Keysight Technologies U1610/20A Service Manual

Handheld digital oscilloscope
Hide thumbs Also See for U1610/20A:
Table of Contents

Advertisement

1
Performance Verification Tests

Multimeter Performance Verification Tests

Use the performance verification tests described in this section to verify the
multimeter measurement performance of the U1610/20A. These tests are based
on the U1610/20A specifications as listed in the U1610/20A User's Guide.
Refer to
tests.
Some of the verification tests involve high voltages; hence, only qualified
WARNING
personnel is recommended to perform the tests.
To avoid electrical shock or personal injury, always place the calibrator in
Standby (STBY) mode before starting the verification test.

DC voltage verification test

1 Put the calibrator in Standby (STBY) mode.
2 On the U1610/20A, press
to select the V DC measurement function.
3 Connect the calibrator to the U1610/20A meter terminals using a banana plug
as shown below. Ensure the polarity of the banana plug connection is correct.
U1610/20A
26
All manuals and user guides at all-guides.com
Table 1-2
for the equipment required for the performance verification
>
(F1) and use the
Fluke 5520A calibrator
Keysight U1610/20A Service Guide
keys