Resistance Verification Test - Keysight Technologies U1610/20A Service Manual

Handheld digital oscilloscope
Hide thumbs Also See for U1610/20A:
Table of Contents

Advertisement

Resistance verification test

1 Put the calibrator in Standby (STBY) mode.
2 On the U1610/20A, press
to select the resistance measurement function.
3 Connect the calibrator to the U1610/20A meter terminals using a 2-wire
compensation mode as shown below. Ensure the polarity of the banana plug
connection is correct.
4 On both the calibrator and the U1610/20A, set each resistance range in the
order shown in
5 Compare the measurement results to the corresponding test limits of
Table
Keysight U1610/20A Service Guide
All manuals and user guides at all-guides.com
U1610/20A
Table
1-6.
1-6.
Performance Verification Tests
>
(F1) and use the
Fluke 5520A calibrator
1
keys
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1610aU1620a