1

Performance Verification Tests

Introduction

The performance verification tests described in this section are used to verify the
measurement performance of the U1610/20A. These tests should be performed
periodically to verify that the U1610/20A is operating within specifications (as
listed in the U1610/20A User's Guide).
If the U1610/20A fails the performance verification tests, adjustment or repair is
required.

Test interval

The recommended interval of the performance verification tests is once per year.
Performance should also be tested after repairs or major upgrades.

Test record

You can use the test record form provided in
Digital Oscilloscope Test Record"
results.

Test consideration

For accurate test results, let the U1610/20A and test equipment warm up for
30 minutes and perform self-calibration before testing. For more information on
performing self-calibration, refer to the U1610/20A User's Guide.
If performance tests fail after self-calibration, send the U1610/20A to your
NOTE
nearest Keysight Service Center for servicing.