Keysight U1610/20A Handheld Digital Oscilloscope Test Record; Oscilloscope Performance Verification Tests - Keysight Technologies U1610/20A Service Manual

Handheld digital oscilloscope
Hide thumbs Also See for U1610/20A:
Table of Contents

Advertisement

1
Performance Verification Tests

Keysight U1610/20A Handheld Digital Oscilloscope Test Record

Serial No.:
Test Interval:
Recommended Next Testing:

Oscilloscope performance verification tests

Voltage measurement accuracy verification test
U1610/20A setting
Calibrator setting
50 V/div
175 V
50 V/div
–175 V
20 V/div
140 V
10 V/div
70 V
5 V/div
35 V
2 V/div
14 V
1 V/div
7 V
500 mV/div
3.5 V
200 mV/div
1.4 V
100 mV/div
700 mV
50 mV/div
350 mV
20 mV/div
140 mV
10 mV/div
70 mV
5 mV/div
35 mV
2 mV/div
14 mV
32
All manuals and user guides at all-guides.com
Minimum test limit
157.40 V
–192.60 V
132.96 V
66.48 V
33.24 V
13.30 V
6.65 V
3.32 V
1.33 V
664.80 mV
332.40 mV
132.96 mV
66.48 mV
33.24 mV
13.30 mV
Test By:
Work Order No.:
Temperature:
Maximum test limit
Channel 1
192.60 V
–157.40 V
147.04 V
73.52 V
36.76 V
14.70 V
7.35 V
3.68 V
1.47 V
735.20 mV
367.60 mV
147.04 mV
73.52 mV
36.76 mV
14.70 mV
Keysight U1610/20A Service Guide
Channel 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1610aU1620a

Table of Contents