Resistance verification test
Calibrator setting
U1610/20A setting
1000 Ω
1000 Ω
10 kΩ
10 kΩ
100 kΩ
100 kΩ
1 MΩ
1 MΩ
10 MΩ
10 MΩ
100 MΩ
100 MΩ
Capacitance verification test
Calibrator setting
U1610/20A setting
1000 nF
1000 nF
10 μF
10 μF
100 μF
100 μF
1000 μF
1000 μF
10 mF
10 mF
Keysight U1610/20A Service Guide
All manuals and user guides at all-guides.com
Minimum limit
996.7 Ω
9.967 kΩ
99.67 kΩ
0.994 MΩ
9.917 MΩ
98.47 MΩ
Minimum limit
987.6 nF
9.876 μF
98.76 μF
979.6 μF
9.796 mF
Performance Verification Tests
Maximum limit
Resul t
1003.3 Ω
10.033 kΩ
100.33 kΩ
1.006 MΩ
10.083 MΩ
101.53 MΩ
Maximum limit
Resul t
1012.4 nF
10.124 μF
101.24 μF
1020.4 μF
10.204 mF