Bandwidth verification test

This test verifies the bandwidth response of the U1610/20A. In this test, you will
use an oscilloscope calibrator to output a leveled sine wave.
You will use the RMS voltage at 1 MHz and at bandwidth frequency to verify the
bandwidth response of the U1610/20A.
From specifications:
U1610A: Bandwidth (±3 dB) = 100 MHz
U1620A: Bandwidth (±3 dB) = 200 MHz
Refer to
Perform the following procedure for channel 1, and then repeat the same
procedure for channel 2.
1 On the U1610/20A, press
channel 1.
2 Connect the calibrator output through a 50 Ω feedthrough termination to the
U1610/20A channel 1 input as shown below.
Keysight U1610/20A Service Guide
All manuals and user guides at all-guides.com
Table 1-1
for the equipment required for this bandwidth verification test.
Fluke 5820A calibrator
Performance Verification Tests
>
>