4 On both the calibrator and the U1610/20A, set each capacitance range in the
order shown in
5 Compare the measurement results to the corresponding test limits of
Table
Table 1-7
Settings for the capacitance verification test
Calibrator setting
Function
1000 nF
10 μF
100 μF
Capacitance
1000 μF
10 mF
Keysight U1610/20A Service Guide
All manuals and user guides at all-guides.com
Table
1-7.
1-7.
U1610/20A setting
1000 nF
10 μF
100 μF
1000 μF
10 mF
Performance Verification Tests
Minimum limit
Maximum limit
987.6 nF
1012.4 nF
9.876 μF
10.124 μF
98.76 μF
101.24 μF
979.6 μF
1020.4 μF
9.796 mF
10.204 mF