Xylem MJK MagFlux Series Manual

Xylem MJK MagFlux Series Manual

Electromagnetic flowmeter
Table of Contents

Advertisement

Quick Links

Manual
EN 3.05 Version 2102
MagFlux® Series
ELECTROMAGNETIC FLOWMETER

Advertisement

Table of Contents
loading

Summary of Contents for Xylem MJK MagFlux Series

  • Page 1 Manual EN 3.05 Version 2102 MagFlux® Series ELECTROMAGNETIC FLOWMETER...
  • Page 2 MagFlux® Series EN 3.05 Version 2102 Your notes:...
  • Page 3: Table Of Contents

    MagFlux® Series EN 3.05 Version 2102 Table of Contents MJK declaration of conformity....................8 Introduction ..........................9 Introduction - MagFlux® Q ....................9 Operating principles ......................9 The MagFlux® Flow Meter....................10 Safety, repair and product identification ................11 Safety instructions........................ 11 Physical mounting........................ 12 Repair ............................
  • Page 4 MagFlux® Series EN 3.05 Version 2102 Connection board - remotely mounted converter............44 System configuration examples....................45 Compact converter and display unit on flow sensor............45 Remote converter with connection box on flow sensor ..........46 Wiring schematic #1......................47 Remote display and multiple converters wiring .............. 48 Wiring schematic #2......................
  • Page 5 MagFlux® Series EN 3.05 Version 2102 Converter software ver..................... 106 Product info ....................... 107 Reset counter time....................107 Internal meas. & cal....................108 Minimum velocity...................... 108 Calibrate mA ......................109 Test velocity....................... 111 Freeze coil ......................... 112 Read eventlog ......................113 Back on stock ......................
  • Page 6 MagFlux® Series EN 3.05 Version 2102 Appendix F. Log files ......................155 Appendix G. Digital input/output connections ..............159 Appendix H. Remote slave display unit ................160 Connection board for remote slave display unit ............161 Converter connection board with master display unit ..........162 Wiring..........................
  • Page 7 MagFlux® Series EN 3.05 Version 2102 Contact You can always contact your local representative or the MJK hotline for advice and guidance: Europe: Tel.: +45 45 56 06 56, e-mail: mjk@mjk.com Visit our web site www.mjk.com to learn more about MJK Automation, our other products and the people behind them.
  • Page 8: Mjk Declaration Of Conformity

    MagFlux® Series EN 3.05 Version 2102 MJK declaration of conformity MJK declaration of conformity /...
  • Page 9: Introduction

    MagFlux® Series EN 3.05 Version 2102 Introduction The MagFlux® is suitable for flow measurement in all kinds of conductive fluids, and it is especially suited for flow measurement of water, waste water, sludge and other fluids containing particles. The flow meter is easy to install and put into service. However, read this manual first to learn about the MagFlux®...
  • Page 10: The Magflux® Flow Meter

    MagFlux® Series EN 3.05 Version 2102 V is the fluid velocity k is a dimensionless constant The flow of the fluid Q (m /s) is given by the following formula: Q = π x D2 x V / 4 => V = 4 / (π x D2) where π...
  • Page 11: Safety, Repair And Product Identification

    MagFlux® Series EN 3.05 Version 2102 Safety, repair and product identification Safety instructions Read this manual carefully. Pay attention to the environment on the installation site. Wear necessary protective equipment and follow all current safety regulations. The MagFlux® can invoke a start signal for dangerous machinery. Always ensure that connected machinery and other equipment are effectively put out of service (that is to remove the main fuses and lock main and security switches in off-position) before commencing configuration, fault...
  • Page 12: Physical Mounting

    MagFlux® Series EN 3.05 Version 2102 115 VAC 230 VAC Physical mounting The MagFlux® flow converter/flow meter must not be mounted in explosion hazardous areas! Repair Repair must only be made by this company or by a service representative approved by us. MJK product identification A delivery will usually consist of a MagFlux®...
  • Page 13 MagFlux® Series EN 3.05 Version 2102 This calibration code (Cal. code) is unique and provides the MagFlux® converter with information about flow sensor number, nominal diameter of the flow sensor and calibration data for the flow sensor. The current converter firmware requires an eight character input, but also accepts six characters plus two OK.
  • Page 14: Flow Sensor

    MagFlux® Series EN 3.05 Version 2102 Flow sensor The following conditions must be obtained to get the full benefit of the MagFlux® flow sensor. Minimum conductivity The conductivity of the measured media must be greater than 5 µS/cm. Liner selection •...
  • Page 15 MagFlux® Series EN 3.05 Version 2102 • Avoid air at the measuring point. • Take care that condensate and water cannot enter the connector box on the flow sensor. • There must be sufficient free space around the flow sensor. However, the MagFlux®...
  • Page 16: Pressure Loss

    MagFlux® Series EN 3.05 Version 2102 Pressure loss Pressure loss can easily be determined, if the nominal pipe diameter is greater than the MagFlux® flow sensor. See this pressure loss diagram: The diagram illustrates that decreasing the internal diameter from 100 mm (DN) to 80 mm (DO) will cause a pressure loss of 0.003 Bar @3 m/s.
  • Page 17: Accuracy

    MagFlux® Series EN 3.05 Version 2102 Accuracy According to the type and size of the flow sensor, the measuring accuracy will be better than 0.25%, provided that the flow sensor has the correct dimension. MJK sizing The flow sensor should be selected so the flow velocity through the sensor will be between 0.2-10 m/s.
  • Page 18 MagFlux® Series EN 3.05 Version 2102 Min and max flow, metrics Qmin Qmax 0.2 m/s 10 m/s 2,39 3,62 5,65 8,84 12,7 22.6 1131 35.3 1767 50.9 2545 69.3 3464 90.5 4524 5726 7069 10179 13854 18095 22902 1000 28274 1200 40715 The Q...
  • Page 19 MagFlux® Series EN 3.05 Version 2102 Min and max flow, imperial Qmin Qmax 0.6 ft./s  30 ft./s [GPM] [GPM] ½” 0.559 28.0 ¾" 0.995 49.76 1” 1.550 77.82 1¼" 2.549 127.4 1½” 3.984 199.7 2” 6.226 310.7 2½” 10.52 523.9 3”...
  • Page 20: Sizing Chart

    MagFlux® Series EN 3.05 Version 2102 Sizing chart Example: A MagFlux® with an internal diameter of 100 mm can measure flow rates between approx. 5.6 m /h and 290 m /h and the fluid velocity is 1.5 m/s at a flow rate of approx. 42 m Flow sensor /Sizing chart...
  • Page 21 MagFlux® Series EN 3.05 Version 2102 Flow sensor /Sizing chart...
  • Page 22 MagFlux® Series EN 3.05 Version 2102 Flow sensor /Sizing chart...
  • Page 23: Pipe System

    MagFlux® Series EN 3.05 Version 2102 Pipe system To ensure a laminar flow without turbulence upstream of the MagFlux® flow sensor, the flow sensor must be mounted in a location which is free from interfering elements like valves, Ts, bends, pumps, etc. For that reason, the MagFlux®...
  • Page 24 MagFlux® Series EN 3.05 Version 2102 pipe installation, it may be necessary with a longer respect distance than stated in the data sheet. For the meter to comply with its measuring accuracy, these respect distances apply both before and after the meter. If in doubt about the installation and the respective respect distances, please contact MJK support hotline...
  • Page 25 MagFlux® Series EN 3.05 Version 2102 The MagFlux® flow sensor should always be filled with liquid. For that reason, the flow sensor must not be mounted at the highest point of the pipe system or in free outlets, where gravity could empty or partially empty the pipe.
  • Page 26 MagFlux® Series EN 3.05 Version 2102 6. When mounting horizontally in pipes with free downstream outlet, the MagFlux® flow sensor should be mounted such that it will always be filled with liquid, for example in a bend situated lower than the height of the outlet.
  • Page 27 MagFlux® Series EN 3.05 Version 2102 7. When mounting horizontally the MagFlux® flow sensor can be rotated max. +/- 45° seen from the connection end. If the flow sensor is rotated more than 45°, one of the electrodes may not be in full contact with the liquid.
  • Page 28: Pipe System - Magflux Q Special Conditions

    MagFlux® Series EN 3.05 Version 2102 Pipe system - MagFlux Q special conditions For MagFlux® Q, a special design allows using a short build-in construction which provides a very accurate flow measurement even at minimum one pipe diameter of straight pipe upstream and one pipe diameter of straight pipe downstream from the centre of the flow sensor.
  • Page 29: Potential Equalization And Grounding

    MagFlux® Series EN 3.05 Version 2102 Potential equalization and grounding Type 7100/7200/8200 in conductive pipes Note! The MagFlux® flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm Note! Does not apply to the MagFlux® Q. MagFlux®...
  • Page 30: Type 7100/7200/8200 In Non-Conductive Pipes

    MagFlux® Series EN 3.05 Version 2102 Type 7100/7200/8200 in non-conductive pipes Note! The MagFlux® flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm Note! Does not apply to the MagFlux® Q. MagFlux®...
  • Page 31: Type 7300 In Conductive Pipes

    MagFlux® Series EN 3.05 Version 2102 Type 7300 in conductive pipes Note! The MagFlux® flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm Type 7300 in non-conductive pipes Note! The MagFlux® flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm Flow sensor /Potential equalization and grounding...
  • Page 32: Cathodic Protection

    MagFlux® Series EN 3.05 Version 2102 Cathodic protection Type 7100/7200 in conductive pipes Note! The Sensor must be isolated from the cathodic protected pipes, this is done by using insulated bolts and nylon washers. Type 7100/7200 in non-conductive pipes Flow sensor /Cathodic protection...
  • Page 33 MagFlux® Series EN 3.05 Version 2102 Note! The Sensor must be isolated from the cathodic protected pipes, this is done by using insulated bolts and nylon washers. Flow sensor /Cathodic protection...
  • Page 34: Flow Converter

    MagFlux® Series EN 3.05 Version 2102 Flow converter Electrical mounting Warning: The MagFlux® flow converter/flow sensor must not be mounted in explosion hazardous areas! Loosen the four screws (position indicated by arrows) and remove the display unit to gain access to the terminals. Flow converter /Electrical mounting...
  • Page 35: Power Supply

    MagFlux® Series EN 3.05 Version 2102 Power supply The MagFlux® flow converter must be supplied from a properly fused mains outlet, a 24 volt AC outlet, or a 10 - 30 V DC power supply/battery. Power Supply 230 V AC, 115 V AC or 24 V AC Terminal Designation Protective ground...
  • Page 36: Change Power Supply Voltage 230/115 V Ac

    MagFlux® Series EN 3.05 Version 2102 Technical Specifications for 10 - 30 V DC Power Supply Power consumption without display < 5 W Power consumption with display < 8 W Peak start current @12 V DC,1 second Approx. 1,5 A Peak start current @24 V DC,1 second Approx.
  • Page 37 MagFlux® Series EN 3.05 Version 2102 6. Unsolder and remove the jumper indicated by the arrow. Flow converter /Change power supply voltage 230/115 V AC...
  • Page 38 MagFlux® Series EN 3.05 Version 2102 Look at the following picture to see where two new jumpers must be inserted (115 V AC configuration): Insert and solder two jumpers (wires) in the positions indicated by the arrows. Turn around the PCB and replace the 63 mAT fuse on the right side with a 125 mAT fuse (or vice verse going from 115 VAC to 230 VAC).
  • Page 39: Analogue Output

    MagFlux® Series EN 3.05 Version 2102 Analogue output The analogue output is an active output with a max. load of 600Ω. Analogue Output Terminal Designation 4-20 mA 4-20 mA The analogue output can be programmed for indication of: • Flow forward •...
  • Page 40: Digital Input

    MagFlux® Series EN 3.05 Version 2102 Mechanical Relay (DO2) Terminal Designation DO 2 Max. 28 VAC/ 28 VDC /500 mA DO 2 DO 1 shares the common terminal (Com) with DI. See also application examples in Appendix G. Digital input/output connections  (see page 159) Digital input...
  • Page 41: Local (Compact) Flow Sensor

    MagFlux® Series EN 3.05 Version 2102 Local (compact) flow sensor Connect the compact (local) flow sensor to the flow converter with the wires coming from the flow sensor as shown below. Compact Flow Sensor Terminal Signal Designation Colour from sensor Built-in liquid ground Liquid GND Black/shield...
  • Page 42 MagFlux® Series EN 3.05 Version 2102 Remote Flow Sensor  691080 Terminal Signal Desig- Color before Color from nation March 2011 March 2011 Liquid Shield 1  Wire pair no. 1 Wire pair White White no. 1 Wire pair Black Black no. 3 Wire pair Green Blue...
  • Page 43: Converter Connection Board - Local

    MagFlux® Series EN 3.05 Version 2102 Converter connection board - local Flow converter /Converter connection board - local...
  • Page 44: Connection Board - Remotely Mounted Converter

    MagFlux® Series EN 3.05 Version 2102 Connection board - remotely mounted converter Note! Before March 2011, the blue wire for terminal 5 was green and the brown wire for terminal 6 was white. Note! After November 2018, the latch between terminal A, 1 and 4 is build into the PCB.
  • Page 45: System Configuration Examples

    MagFlux® Series EN 3.05 Version 2102 System configuration examples Compact converter and display unit on flow sensor Configuration: The MagFlux® converter and display unit is mounted directly on the MagFlux® flow sensor. Order numbers for this configuration: • 207xxx MagFlux® Flow Sensor •...
  • Page 46: Remote Converter With Connection Box On Flow Sensor

    MagFlux® Series EN 3.05 Version 2102 Remote converter with connection box on flow sensor Configuration: The MagFlux® converter and display unit is remote mounted for example when the sensor is to be buried or submerged. Order numbers for this configuration: •...
  • Page 47: Wiring Schematic #1

    MagFlux® Series EN 3.05 Version 2102 Wiring schematic #1 Wiring schematic for Remote converter with connection box on flow sensor  (see page 46) Flow sensor Connect one end of the cable to the flow sensor as described in the table below. All three shields are twisted and connected in slot 1.
  • Page 48: Remote Display And Multiple Converters Wiring

    MagFlux® Series EN 3.05 Version 2102 Remote display and multiple converters wiring Configuration: Two locally mounted MagFlux® flow sensors and converters with blind lid and one remotely mounted display unit. The communication between the sensors and the display unit is executed on shielded twisted 2- and 4-wire cables using the Modbus®...
  • Page 49 MagFlux® Series EN 3.05 Version 2102 Local flow sensor 1 2-wire twisted shield cable: Connect lead 1 to slot A, lead 2 to slot B and shield to slot -. Local flow sensor 2 2-wire twisted shield cable: Connect lead 1 to slot A, lead 2 to slot B and twisted shield to slot -.
  • Page 50: Startup

    MagFlux® Series EN 3.05 Version 2102 Startup Initial checks Before switching power on for the MagFlux®, the following steps must be checked. The local mains power supply voltage corresponds to the voltage printed on the identification label of the MagFlux® flow converter. All electrical connections are made in accordance with the electrical connection diagram shown in Converter connection board - remote...
  • Page 51: Language Selection

    MagFlux® Series EN 3.05 Version 2102 Language selection The default display language is English (GB). If another language is required, proceed with step 2. From the Main Menu, select Display Setup  (see page 116) Select Language and chose the required language.  (see page 117) Display read-out, one connected unit All MagFlux®...
  • Page 52: Display Read-Out, Several Connected Units

    MagFlux® Series EN 3.05 Version 2102 to appear like this: Display read-out, several connected units When several units are interconnected, for example an Oxix® dissolved oxygen transmitter and MagFlux® flow meter with different names and Modbus ID numbers, a Display Overview Menu is available at top level (press Back repeatedly): All connected units are displayed and sorted by their ID numbers, and consequently each unit can be selected and set up as required:...
  • Page 53: Display Keys

    MagFlux® Series EN 3.05 Version 2102 Important: More connected units can only by managed as described above if each unit has been assigned a unique name and Modbus ID number. See Modbus for details.  (see page 119) Display keys The keys and the soft keys (variable key functions determined by the display firmware) are used for initial programming and normal operation of the flow meter.
  • Page 54: Initial Setup

    MagFlux® Series EN 3.05 Version 2102 Save the new setting by pressing the two outmost keys simultaneously. System reset You can reset and refresh all system displays and key combinations by pressing all four keys simultaneously. This feature is especially useful during for example a service call, where the display language (Dutch, Danish, etc.) cannot be understood by the service person.
  • Page 55 MagFlux® Series EN 3.05 Version 2102 If changes to the unique sensor calibration code are needed following physical installation and initial setup, the Sensor Calibration Code menu must be addressed. A password is not required to enter or change the sensor calibration code.
  • Page 56 MagFlux® Series EN 3.05 Version 2102 Select Service Menu, enter the password (the default password is "00000" (five zeros) in Sensor Calibration Code. Enter the calibration code read from the flow sensor label (or pass through or change the displayed calibration code) and press OK. See also details regarding the calibration code in the section on product identification.
  • Page 57: Magflux® Menus

    MagFlux® Series EN 3.05 Version 2102 MagFlux® menus All the MagFlux® menus and sub menus are shown and described in the following sections. Continuous overviews of the menu and sub-menu structures are available from the appendix. Main menu Press the Setup key on the MagFlux® display (see below) to enter the MagFlux®...
  • Page 58: Specify Main Screen

    MagFlux® Series EN 3.05 Version 2102 Specify main screen The Specify Main Screen menu allows you to customize the MagFlux® display to suit your requirements. You can add and remove the five available display lines and configure them individually. MagFlux® menus /Specify main screen...
  • Page 59 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Specify main screen...
  • Page 60 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Specify main screen...
  • Page 61 MagFlux® Series EN 3.05 Version 2102 Press the up/down keys to highlight the required menu line and press Press the up/down keys to highlight the required option and press OK. The available options are: • Not in use: The line will not be used. The set free space will be used by the other lines.
  • Page 62: Factory Settings

    MagFlux® Series EN 3.05 Version 2102 The size of the display lines will automatically increase or decrease as the number of display lines is removed or added to maximize the field of view for the measured values. Factory settings The Factory Settings menu resets the display to default settings, to metric settings or to US settings.
  • Page 63 MagFlux® Series EN 3.05 Version 2102 The available options are: • Cancel: Exit the menu without changes. • Default: When Default is selected, the following settings are not affected and remain as chosen by the user: • Sensor Name • Device ID •...
  • Page 64 MagFlux® Series EN 3.05 Version 2102 Per default, the in- and outputs are not activated from the factory. Activate the in- and outputs by configuring the functions with, for example, the default values shown in the following tables: • MagFlux Default DI/DO Settings •...
  • Page 65 MagFlux® Series EN 3.05 Version 2102 MagFlux Default 20 mA Settings Metric DN [mm] Size Flow [GPM] Flow [m ½" 3/4" 1" 1¼" 1½" 2" 2½" 3" 4" 1000 5" 2000 6" 3000 1000 8" 5000 2000 10" 7700 2500 12"...
  • Page 66: Data Logger

    MagFlux® Series EN 3.05 Version 2102 • Set to metric: All converter-related units can be set to the following units (default values in bold and italics): Flow Tota- Unit lizers RFTot RRTot RSTot FTot RTot /sec l/sec l/min • Set to US: All converter-related units can be set to the following units (default values bold and in italics) Flow Tota-...
  • Page 67 MagFlux® Series EN 3.05 Version 2102 Select the digits one by one with the left/right keys and set the values with the up/down keys. Continue until all digits have been set and press OK. The log interval can be set in intervals from 10 sec. up to 9999 sec. The data log contains: •...
  • Page 68: Graph Display

    MagFlux® Series EN 3.05 Version 2102 read in, for example Microsoft® Excel®, (see Appendix B. about Field Link software for details). An example of the log capacity of one sensor versus the time interval is shown in the following table. Log interval Log capacity Seconds...
  • Page 69: Password

    MagFlux® Series EN 3.05 Version 2102 The MagFlux® display unit then shows the Graph screen. To return to the Main Menu, select esc by pressing the up/down keys simultaneously. The Y axis is automatically scaled according to the Q of the mA output. Double-arrow keys jump forward and backward one screen frame at a time.
  • Page 70: Set Sensor Name

    MagFlux® Series EN 3.05 Version 2102 Press the up/down keys to highlight the required menu line and press The available options are: • Login: Select the digits one by one with the left/right keys and set the values with the up/down keys and press OK. •...
  • Page 71 MagFlux® Series EN 3.05 Version 2102 display with up to four display lines. Press the up/down keys to highlight the required menu line (here: Sensor Name) and press OK. MagFlux® menus /Set sensor name...
  • Page 72 MagFlux® Series EN 3.05 Version 2102 To change the default sensor name (MagFlux1), press the left/right keys to highlight the wanted character. Select the digits one by one with the left/right keys and set the value with the up/down keys. Continue until the name has been set to, for example, Inlet flow (see the following).
  • Page 73: Converter Setup

    MagFlux® Series EN 3.05 Version 2102 Converter setup Converter Setup provides configuration options for volumes, batches, units, etc. See detailed descriptions in the following sections and in the overview sections at the end. Press the up/down keys to highlight the required menu line (here: Converter Setup) and press OK.
  • Page 74: Minimum Flow

    MagFlux® Series EN 3.05 Version 2102 2. Press the up/down keys to highlight the required option and press OK. The available options are described in detail in the following sections. Minimum flow Sets the minimum forward and reverse flow rate. MagFlux® flow meters are default set to 0,1 m/s.
  • Page 75 MagFlux® Series EN 3.05 Version 2102 The 1 m/s indicates the flow volume at the selected unit. Select the digits one by one with the left/right keys and set the value with the up/down keys. When you press OK, Reverse and Forward will change places and you can edit the other one.
  • Page 76: Averaging

    MagFlux® Series EN 3.05 Version 2102 Averaging Standard averaging (the first setting) indicates the time period within which measurements are smoothed and averaged. In Averaging, to change the standard averaging period, select the digits one by one with the left/right keys and set the value with the up/down keys.
  • Page 77: Units

    MagFlux® Series EN 3.05 Version 2102 Units Sets the unit for the flow rate. The available units are shown below. Press the up/down keys to highlight the required unit and press OK. mA output When a MagFlux® is connected to a power supply for the first time, the mA output is automatically set to provide 4 mA at zero flow and 20 mA at a flow corresponding to the theoretical Q of the flow sensor.
  • Page 78: Flow Forward

    MagFlux® Series EN 3.05 Version 2102 Using the factory setup, MagFlux® returns to default mA settings corresponding to the chosen MagFlux® flow sensors Q and Q . The mA output is an active output, and the maximum load is 600 Ohm. The upper limit for the mA output is 20,5 mA: •...
  • Page 79: Flow Reverse

    MagFlux® Series EN 3.05 Version 2102 Flow reverse The mA signal provides 4 mA at zero flow and 20 mA at Q in the reverse direction. The 1 m/s indicates the flow volume at the selected unit. MagFlux® menus /Converter setup...
  • Page 80: Forward & Rev. (12Ma=0)

    MagFlux® Series EN 3.05 Version 2102 Forward & rev. (12mA=0) The mA signal provides 4 to 20 mA reverse to forward direction with 12 mA at zero flow. The 1 m/s indicates the flow volume at the selected unit. MagFlux® menus /Converter setup...
  • Page 81: Forward & Rev. (4Ma=0)

    MagFlux® Series EN 3.05 Version 2102 Forward & rev. (4mA=0) The mA signal provides 4 to 20 mA for forward and reverse flow with 4 mA at zero flow. The 1 m/s indicates the flow volume at the selected unit. MagFlux®...
  • Page 82: Not In Use

    MagFlux® Series EN 3.05 Version 2102 Not in use The mA Output option is not being used. The 1 m/s indicates the flow volume at the selected unit. Totalizers The MagFlux® provides six totalizers, each with two or three output options. Up to two totalizers can be assigned to one digital output.
  • Page 83: Settings And Limits For Resettable Totalizers

    MagFlux® Series EN 3.05 Version 2102 • Non-Resettable Totalizer Forward • Non-Resettable Totalizer Reverse • Non-Resettable Totalizer Sum The available output options for the resettable totalizers are: • Mechanical Relay (DO2) • Opto Relay (DO1) (light-triggered electronic relay) • Display only (on the display unit) The available output option for the non-resettable totalizers is: •...
  • Page 84 MagFlux® Series EN 3.05 Version 2102 MagFlux® Totalizer Output Setting Metric Optical DO Mechanical 1 @100 DO 2 @100 mSec. mSec. DN [mm] min. max. min. max. Unit 0,001 0,001 0,01 0,01 0,01 0,01 0,01 1000 1000 1000 1000 1000 1000 1000 1000...
  • Page 85 MagFlux® Series EN 3.05 Version 2102 MagFlux® Totalizer Output Setting Metric Optical DO Mechanical 1 @100 DO 2 @100 mSec. mSec. DN [mm] min. max. min. max. Unit 10000 10000 10000 1000 10000 10000 1000 10000 10000 1000 10000 1000 10000 1000 10000...
  • Page 86: Resettable Totalizers

    MagFlux® Series EN 3.05 Version 2102 The formula for minimum flow unit per pulses is: (flow at 20mA) / 120 Example: DN 100 max. flow 300 m /h ? Minimum flow unit per pulses = 300 / 120 = 2,5 m which is rounded to 10 m by MagFlux®...
  • Page 87 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Converter setup...
  • Page 88 MagFlux® Series EN 3.05 Version 2102 Resettable forward totalizer RFTot: The Totalizer counts the forward flow volume according to the primary flow direction selected at startup. The Totalizer counts in selected units and is resettable. The counter can be connected to the relays or to the display only. To reset the counter, the totalizer must be present in the main display and be selected by the up/down keys followed by a click on Reset.
  • Page 89: Batch Counters 1 & 2

    MagFlux® Series EN 3.05 Version 2102 NR totalizer forward NR FTot: The Totalizer counts the forward flow volume according to the primary flow direction selected at start up. The Totalizer counts in selected units and cannot be reset. The counter can only be selected from the Main display. NR totalizer reverse NR RTot: The Totalizer counts the reverse flow volume according to the primary flow direction selected at start up.
  • Page 90 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Converter setup...
  • Page 91 MagFlux® Series EN 3.05 Version 2102 In the following descriptions of the batch counters, the function or digital status is illustrated as follows: Automatic Batch Counter: The automatic batch counter issues a signal, when a preset volume is reached. On-time: The period of time in which the relay is activated (drawn). Number of preselected (units): The preset batch volume.
  • Page 92 MagFlux® Series EN 3.05 Version 2102 Manual Batch Counter: The counter issues signals as determined by manual start, stop and pause commands. A manual stop is equivalent to a reset command. Number of preselected (units): The preset batch volume. MagFlux® menus /Converter setup...
  • Page 93 MagFlux® Series EN 3.05 Version 2102 Adaptive Batch Counter: The counter compensates for overrun and insufficient flow. Number of preselected (units): The preset batch volume. Maximum Correction: The correction in percent that triggers an error message or an alarm. Reset Batch Error: Reset the batch correction percentage to 0%. MagFlux®...
  • Page 94: Settings And Limits For Batch Counters

    MagFlux® Series EN 3.05 Version 2102 Settings and limits for batch counters • Batch counter units: Available units are l, hl, kl, m , ft , gal and MG. • Start, stop or pause with Digital Input: The manual and the adaptive batch counters can be started, stopped or paused via the digital input (DI) •...
  • Page 95 MagFlux® Series EN 3.05 Version 2102 MagFlux® Batch Counter Output Settings Metric Optical DO1 @100 mSec. Mechanical DO2 @100 mSec. min. max. min. max. Unit [mm] Auto- Manually/ Auto- Manually/ matic adaptive matic adaptive 0,041666 0,0002778 0,0138889 0,0138889 0,083333 0,0005556 0,0277778 0,0277778 0,001111 0,055556...
  • Page 96 MagFlux® Series EN 3.05 Version 2102 MagFlux® Batch Counter Output Settings Metric Optical DO1 @100 mSec. Mechanical DO2 @100 mSec. min. max. min. max. Unit [mm] Auto- Manually/ Auto- Manually/ matic adaptive matic adaptive 0,333333 16,666667 6000 50,00000 16,666667 6000 0,388889 19,444444 7000 58,33333 19,444444...
  • Page 97: High Flow/Low Flow

    MagFlux® Series EN 3.05 Version 2102 Limitation of the automatic batch counter on DO2 (mech. relay): The maximum limit is 120 pulses per hour. The formula for minimum flow unit per pulses is: (flow at 20mA) / 120 Example: DN 100 max. flow 300 m /h ? Minimum flow unit per pulses = 300 / 120 = 2,5 m The formula for maximum flow unit per pulses is: (flow at 20mA)
  • Page 98 MagFlux® Series EN 3.05 Version 2102 The available output options for High Flow and Low Flow are: MagFlux® menus /Converter setup...
  • Page 99: Flow Direction

    MagFlux® Series EN 3.05 Version 2102 • Mechanical Relay (DO2) • Opto Relay (DO1) (light-triggered, electronic relay) • Display only (signals to the display unit only) • Not in use (the High/Low Flow option is not being used) ON at Flow: Issues an alarm and/or sends a signal when, for example, a high flow limit is exceeded.
  • Page 100 MagFlux® Series EN 3.05 Version 2102 The available output options for Flow Direction are: • Mechanical Relay (DO2) • Opto Relay (DO1) (light-triggered electronic relay) • Display only (signals to the display unit only) • Not in use (the Flow Direction option is not being used) Delay: A delay in seconds can be set to compensate for positive and negative spikes in the flow.
  • Page 101: Empty Pipe

    MagFlux® Series EN 3.05 Version 2102 • NC (Normally Closed) determines the relay state under normal conditions. Empty pipe A signal and/or an alarm can be set, when the sensor becomes practically empty (see also Pipe system ), or if the conductivity drops below 5  (see page 23) µS/cm.
  • Page 102: 24H Flow

    MagFlux® Series EN 3.05 Version 2102 The available output options for Empty Pipe are: • Mechanical Relay (DO2) • Opto Relay (DO1) (light-triggered electronic relay) • Display only (signals to the display unit only) • Not in use (the Empty Pipe option is not being used) Delay: A delay in seconds can be set to compensate for positive and negative spikes in the flow.
  • Page 103 MagFlux® Series EN 3.05 Version 2102 • /: A blinking forward slant (/) indicates that the input/output is in the process of being activated. Eventually, it turns into a steady X. • \: A blinking backward slant (\) indicates that the input/output is in the process of being deactivated.
  • Page 104 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Converter setup...
  • Page 105: Service Menu

    MagFlux® Series EN 3.05 Version 2102 The available output options for Status are: • DO1: Status for Digital Output 1. • DO2: Status for Digital Output 2. • DI: Status for Digital Input. • mA output: The present analog output current in mA. Pending errors: Lists pending errors.
  • Page 106: Converter Software Ver

    MagFlux® Series EN 3.05 Version 2102 where to find it. A password is not required to add or change the sensor calibration code. Converter software ver. The converter software version and bootload version are displayed, so that you can determine whether a software (firmware) upgrade is required or not. MagFlux®...
  • Page 107: Product Info

    MagFlux® Series EN 3.05 Version 2102 Product info Key product data like serial number, hardware type no., track no., etc. are displayed. Reset counter time The counter reset and run-time can be reset for fault finding, error correction and similar service procedures. MagFlux®...
  • Page 108: Internal Meas. & Cal

    MagFlux® Series EN 3.05 Version 2102 Internal meas. & cal. For our service technician's use only. Internal measurements and calculations used for adjustments and calibrations. Minimum velocity Sets the minimum forward and reverse flow velocity. MagFlux® flow meters are default set to 0,1 m/s. For other values, see Sizing Chart .
  • Page 109: Calibrate Ma

    MagFlux® Series EN 3.05 Version 2102 Calibrate mA From this menu, the mA reading can be read, set and calibrated against a multimeter. • Current: Indicates the actual current output. • Set fixed current: Lets you set the mA to a fixed output of either 4, 12 or 20 mA or to a minimum mA or a maximum mA, primarily used for test purposes.
  • Page 110 MagFlux® Series EN 3.05 Version 2102 MagFlux® menus /Converter setup...
  • Page 111: Test Velocity

    MagFlux® Series EN 3.05 Version 2102 Test velocity For our service technician's use only. Can be used to compare the actual velocity against the flow/velocity/diameter chart. MagFlux® menus /Converter setup...
  • Page 112: Freeze Coil

    MagFlux® Series EN 3.05 Version 2102 Freeze coil For our service technician's use only. The sensor coil current and voltage values can be "frozen" to enable read-out using an standard multimeter. MagFlux® menus /Converter setup...
  • Page 113: Read Eventlog

    MagFlux® Series EN 3.05 Version 2102 Read eventlog See the events within a certain time frame. Back on stock Resets the factory settings. This should be done by trained personnel only and a password login is required. MagFlux® menus /Converter setup...
  • Page 114: Popup On/Off

    MagFlux® Series EN 3.05 Version 2102 Popup on/off Lets you determine which pop-up alarms you wish to see in your MagFlux® display. Click Change to disable a popup when it is selected.  The following item is for future usage only! Zero point calibration If, for some reason, your MagFlux®...
  • Page 115 MagFlux® Series EN 3.05 Version 2102  The following item is for future usage only! MagFlux® menus /Converter setup...
  • Page 116: Verification

    MagFlux® Series EN 3.05 Version 2102 Verification This part of the menu requires special training and special equipment. Contact your us for assistance. Display setup The Display Setup menu provides configuration options for language, clock and factory settings and Modbus parameters and software version display. Press the up/down keys to highlight the required menu line (here: Display Setup) and press OK.
  • Page 117: Mjk Communication Modules Documentation

    MagFlux® Series EN 3.05 Version 2102 The available options are described in detail on the following sections. Note: The option Modbus Com Module is only visible (and selectable), if a communication module is installed and connected to the MagFlux® display. If the communication module is removed, the option disappears automatically.
  • Page 118 MagFlux® Series EN 3.05 Version 2102 (below languages are examples). Chose the required language using the up/down keys up and press OK. Additional languages can be added using the Field Link software, see Appendix B. MagFlux® menus /Display setup...
  • Page 119: Set Clock

    MagFlux® Series EN 3.05 Version 2102 Set clock The Set Time/Date menu provides setting of the built-in clock and time format. Select whether daylight saving is in use or not. Select the time format. Select the date and time digits with the left/right keys and change the setting with the up/down keys.
  • Page 120: Add Device

    MagFlux® Series EN 3.05 Version 2102 When several converters are to be added, it is necessary to change each converter device address to a unique ID address. This can be done from the display before doing the Modbus connection. Add Device Four MagFlux®...
  • Page 121: Change Device Address

    MagFlux® Series EN 3.05 Version 2102 Change Device Address To change each converter's device address, it is required that you connect the MagFlux® display unit to each converter one by one. If neglected, a conflict between the units will arise when the RS 485 serial loop is established. Change for example Device 1 to Device 2 as follows: Press OK and use the up/down keys to change 1 to 2.
  • Page 122: Remove Device

    MagFlux® Series EN 3.05 Version 2102 The MagFlux® converter ID is now changed to 2, and more converters can be added to the display unit with the Add Device option. Remove Device If a device is to be removed from the display unit, proceed as follows: Select the required device with the up/down keys.
  • Page 123: Change Display Id No

    MagFlux® Series EN 3.05 Version 2102 Change Display ID No. The display ID number and the number of displays may have to be changed to avoid conflicts on the Modbus. Set No. of Retries If the data communication lines are subject to noise or other disturbances, the number of retries can be raised to increase the chances of a successful change of settings.
  • Page 124: Change Modbus Speed

    MagFlux® Series EN 3.05 Version 2102 Change Modbus speed Change between high and low Modbus speed. Default is low. Factory setting setup The Factory Settings menu provides log deletion and resetting the display unit to factory settings. Press the up/down keys to highlight the required option and press OK. Important: The Data Logger will be reset, and all devices will be deleted from the display Unit.
  • Page 125: Display Sw Version

    MagFlux® Series EN 3.05 Version 2102 Display SW version The Display SW Version menu provides a read-out of the display software (actually firmware) version, the build date and last-edited date of the the multiple languages file. MagFlux® menus /Display setup...
  • Page 126: Mechanical Dimensions

    MagFlux® Series EN 3.05 Version 2102 Mechanical dimensions MJK standard flanges - EN-1092-1_2001 - Europe Flange marking: zzz/EN 1092-1/01 A/DN xxxx/PN ww/S235JR/yyyyy where: zzz = Manufacturer's name/trademark EN 1092-1 = Standard number 01 A = Flange type DN xxxx = DN (size) PN ww = PN (pressure) S235JR = Material...
  • Page 127 MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - EN-1092-1:2001  Size Pres-  f sure Outer Bolt Thickness [mm] No. of Hole Bolt Flan- dia- dia- holes dia- size meter meter meter type [mm] [bar] [mm] [mm] PN [mm] Mechanical dimensions /MJK standard flanges - EN-1092-1_2001 - Europe...
  • Page 128: Magflux Q Sensor Sizes And Torque

    MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - EN-1092-1:2001  Size Pres-  f sure Outer Bolt Thickness [mm] No. of Hole Bolt Flan- dia- dia- holes dia- size meter meter meter type [mm] [bar] [mm] [mm] PN [mm] 1075 1020 1000 1175 1120...
  • Page 129: Mjk Standard Flanges - Ansi B 16.5 - U.s.a

    MagFlux® Series EN 3.05 Version 2102 Tighten the nuts to the needed torque in the illustrated order. Tighten in four steps: 100% 100% clockwise from top MJK standard flanges - ANSI B 16.5 - U.S.A.   Mechanical dimensions /MJK standard flanges - ANSI B 16.5 - U.S.A.
  • Page 130 MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - ANSI B 16.5 Size Pressure No. of class holes Outer Bolt Thick- Hole diameter diameter ness diameter [in] [psi] 1) [in] [in] [in] [in] ½" 3.50 2.38 0.44 0.62 3/4" 3.88 2.75 0.50 0.62...
  • Page 131 MagFlux® Series EN 3.05 Version 2102 36" Class D 46.00 42.75 1.63 1.63 40" Class D 50.75 47.25 1.63 1.63 48" Class D 59.50 56.00 1.89 1.63 1) 600-900 psi: Consult the factory. Flange Dimensions - ANSI B 16.5 Size Pressure No.
  • Page 132: Mjk Standard Flanges - As-4087-2004 - Australia

    MagFlux® Series EN 3.05 Version 2102 20" 30.50 27.00 2.44 1.38 24" 36.00 32.00 2.69 1.63 Flange Dimensions - AWWA C207-01 28" Class E 36.50 34.00 2.06 1.38 32" Class E 41.75 38.50 2.25 1.63 36" Class E 46.00 42.75 2.38 1.63 40"...
  • Page 133 MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - PN 16 - AS-4087-2004 Size No. of Outer Bolt Thick- Hole holes diameter diameter ness diameter [mm] [mm] [mm] [mm] [mm] 1060 1175 1092 Mechanical dimensions /MJK standard flanges - AS-4087-2004 - Australia...
  • Page 134: Mjk Standard Flanges - As-4087-2004 (Continued #1)

    MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - PN 16 - AS-4087-2004 Size No. of Outer Bolt Thick- Hole holes diameter diameter ness diameter [mm] [mm] [mm] [mm] [mm] 1000 1255 1175 1200 1490 1410 MJK standard flanges - AS-4087-2004 (continued #1)  ...
  • Page 135 MagFlux® Series EN 3.05 Version 2102 Flange dimensions - PN21 - AS 4087-2004 Size No. of Outer Bolt Thickness Hole holes diameter diameter diameter [mm] [mm] [mm] [mm] [mm] 1060 1185 1105 1000 1275 1194 1200 1530 1441 Mechanical dimensions /MJK standard flanges - AS-4087-2004 (continued #1)
  • Page 136: Mjk Standard Flanges - As-4087-2004 (Continued #2)

    MagFlux® Series EN 3.05 Version 2102 MJK standard flanges - AS-4087-2004 (continued #2)   Mechanical dimensions /MJK standard flanges - AS-4087-2004 (continued #2)
  • Page 137 MagFlux® Series EN 3.05 Version 2102 Flange dimensions - PN35 - AS-4087-2004 Size No. of Outer Bolt Thickness Hole holes diameter diameter diameter [mm] [mm] [mm] [mm] [mm] 1060 1185 1105 1000 1275 1194 1200 1530 1441 Mechanical dimensions /MJK standard flanges - AS-4087-2004 (continued #2)
  • Page 138: Flygt Standard Flanges - En-1092-1_2001 - Europe

    MagFlux® Series EN 3.05 Version 2102 Flygt standard flanges - EN-1092-1_2001 - Europe Flange marking: zzz/EN 1092-1/01 A/DN xxxx/PN ww/S235JR/yyyyy where: zzz = Manufacturer's name/trademark EN 1092-1 = Standard number 01 A = Flange type DN xxxx = DN (size) PN ww = PN (pressure) S235JR = Material...
  • Page 139 MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - EN-1092-1:2001  Size Pres-  f sure Outer Bolt Thickness [mm] No. of Hole Bolt Flan- dia- dia- holes dia- size meter meter meter type [mm] [bar] [mm] [mm] PN [mm] 1075 1020 Mechanical dimensions /Flygt standard flanges - EN-1092-1_2001 - Europe...
  • Page 140 MagFlux® Series EN 3.05 Version 2102 Flange Dimensions - EN-1092-1:2001  Size Pres-  f sure Outer Bolt Thickness [mm] No. of Hole Bolt Flan- dia- dia- holes dia- size meter meter meter type [mm] [bar] [mm] [mm] PN [mm] 1000 1175 1120 1200 1405...
  • Page 141: Appendix A. Pop-Up And Error Messages

    MagFlux® Series EN 3.05 Version 2102 Appendix A. Pop-up and error messages Alarms The converter has experienced HW problems measuring flow (see Sensor error log for details/time) Sensor cal. code The calibration code conflicts with the sensor HW error The converter has experienced problems with connection to the Output coil error sensor coil (see log for details/time) Empty pipe...
  • Page 142 MagFlux® Series EN 3.05 Version 2102 Alarms Flow is over or under settings for current (it can be that flow is to mA flow error high or negative) Batch1 adaptive Batch 1 adaptive flow is over/under the limit for error correction error Batch2 adaptive Batch 2 adaptive flow is over/under the limit for error correction...
  • Page 143 MagFlux® Series EN 3.05 Version 2102 Exceptions Used for resettable The relay is already in use by another function. Select a different totalizer total relay Used for batch The relay is already in use by another function. Select a different counter 1 relay Used for batch...
  • Page 144: Appendix B. Mjk Field Link™ Software

    MagFlux® Series EN 3.05 Version 2102 Appendix B. MJK Field Link™ software The MJK Field Link™ software package provides several utilities that is described in detail in the manual regarding MJK Field Link™ which is embedded in the software under Help. Important: Current MagFlux®...
  • Page 145 MagFlux® Series EN 3.05 Version 2102 Unscrew the four screws that hold the display unit. Lift out the display unit and connect a USB mini A/B cable to the mini USB female B connector on the rear of the front panel. Connect the other end of the USB cable (max.
  • Page 146: Save Log Data

    MagFlux® Series EN 3.05 Version 2102 Save log data Select LogData in the menu bar and click Save Log to File. Assign a file name, find the destination you want to save the file to, and click Save. Consequently, the file is saved, and the contents is displayed.
  • Page 147: Upgrade Converter Firmware

    MagFlux® Series EN 3.05 Version 2102  The display unit firmware is then transferred, and the display unit is consequently programmed. Expect this process to take a few minutes.  A Searching for Field dialogue box then appears and prompts for disconnecting and re-connecting the USB cable.  Do as suggested and click OK.
  • Page 148: Install/Add Languages

    MagFlux® Series EN 3.05 Version 2102 process to take a few minutes.   When the Flash programming is done, click OK on the PC and the MagFlux® Display Unit to return to the MagFlux® Flow/FTot display.  If required you may adjust the MagFlux® Display Unit screen contrast by pressing the two outmost keys simultaneously, press the up/down keys as required, and save the new setting by again pressing the two outmost keys simultaneously.
  • Page 149 MagFlux® Series EN 3.05 Version 2102 5. The languages are then transferred and loaded. If required, you can remove one or more languages by selecting them from the list, clicking Delete and Close. Appendix B. MJK Field Link™ software /Install/Add languages...
  • Page 150: Appendix C. Faqs

    MagFlux® Series EN 3.05 Version 2102 Appendix C. FAQs Question: Which size MagFlux® flow meter should I choose for my installation? I would prefer a flow meter with the same diameter as the tube(s). Answer: The minimum and maximum flows determine the size. Use the Sizing chart to find the correct flow meter size, and avoid selecting a too...
  • Page 151 MagFlux® Series EN 3.05 Version 2102 Question: I have an "Empty pipe" indication, but the pipe is not empty! Answer: There are no grounding rings, or the rings have not been properly grounded. See examples in Potential equalization and grounding  (see page 29) Question: I can't set the counter to the required output, only to a value that is higher.
  • Page 152: Appendix D. Front Panel Cut-Out Drawing

    MagFlux® Series EN 3.05 Version 2102 Appendix D. Front panel cut-out drawing The below is a representation of a cut-out drawing for the MagFlux® front panel. The dotted line indicates the front panel outline and measures 155 x 145 mm or 6.10 x 5.71 inches.
  • Page 153: Mjk Cut-Out Download

    MagFlux® Series EN 3.05 Version 2102 MJK cut-out download The real-size drawing can be downloaded for A4 and letter format here: Download Center https://mjk.com/download-center?mainCatId=3103&catId=3266&lang=english Appendix D. Front panel cut-out drawing /MJK cut-out download...
  • Page 154: Appendix E. Mjk Test Certificate

    MagFlux® Series EN 3.05 Version 2102 Appendix E. MJK test certificate Appendix E. MJK test certificate /MJK cut-out download...
  • Page 155: Appendix F. Log Files

    MagFlux® Series EN 3.05 Version 2102 Appendix F. Log files The example log file shown below is the result of a CSV file from the converter having been opened in with Microsoft® Excel® spread sheet utility. This example in the following illustrates four MagFlux® converters being logged every 10 secs.
  • Page 156 MagFlux® Series EN 3.05 Version 2102 Appendix F. Log files /MJK cut-out download...
  • Page 157 MagFlux® Series EN 3.05 Version 2102 The entries are described on the following section. • Index: Converter ID (here 4 converters: 32, 33, 34 and 35) • MB Address: Modbus address (here: 1, 2, 3 and 4) • Unit: SI unit according to the unit codes (here: 19 = m /h) ...
  • Page 158 MagFlux® Series EN 3.05 Version 2102 • Column J: The sensor name as entered by the operator. • Time: Date and time in Central European Standard Time (CEST) • UTime: The time in UNIX format. Number of seconds since January 1, 1970 •...
  • Page 159: Appendix G. Digital Input/Output Connections

    MagFlux® Series EN 3.05 Version 2102 Appendix G. Digital input/output connections The digital in- and outputs, DI and DO, can be interconnected to external equipment to achieve a number of functions such as alarms, counter reset, flow direction indication, etc. The two schematics below illustrate a simplified mode of operation.
  • Page 160: Appendix H. Remote Slave Display Unit

    MagFlux® Series EN 3.05 Version 2102 Appendix H. Remote slave display unit This example describes the cable connections and display configurations for one MagFlux® converter with a local "master" display unit and one remote "slave" display unit. Appendix H. Remote slave display unit /MJK cut-out download...
  • Page 161: Connection Board For Remote Slave Display Unit

    MagFlux® Series EN 3.05 Version 2102 Connection board for remote slave display unit Appendix H. Remote slave display unit /Connection board for remote slave display unit...
  • Page 162: Converter Connection Board With Master Display Unit

    MagFlux® Series EN 3.05 Version 2102 Converter connection board with master display unit Appendix H. Remote slave display unit /Converter connection board with master display unit...
  • Page 163: Wiring

    MagFlux® Series EN 3.05 Version 2102 Wiring Connect a lead between: Display, A on the Converter Connection Board to Connection for remote display unit, A on the Connection Board. Connect a lead between: Display, B on the Converter Connection Board to Connection for remote display unit, B on the Connection Board.
  • Page 164: Main Menu Overview

    MagFlux® Series EN 3.05 Version 2102 Main menu overview The size of this manual does not provide sufficient space for showing the complete, contiguous menu structure. As a compensation, a cut-up presentation of the main-menu structure is shown here and relevant sub-menus in the following sections.
  • Page 165 MagFlux® Series EN 3.05 Version 2102 Main menu overview /Configuration for the slave unit...
  • Page 166: Converter Setup Menu Overview

    MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview The size of this manual does not provide sufficient space for showing the complete, contiguous menu structure. As a compensation a cut-up presentation of the menu structure is shown. Converter setup menu overview /Configuration for the slave unit...
  • Page 167 MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview /Configuration for the slave unit...
  • Page 168 MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview /Configuration for the slave unit...
  • Page 169 MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview /Configuration for the slave unit...
  • Page 170 MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview /Configuration for the slave unit...
  • Page 171 MagFlux® Series EN 3.05 Version 2102 Converter setup menu overview /Configuration for the slave unit...
  • Page 172: Service Menu Overview

    MagFlux® Series EN 3.05 Version 2102 Service menu overview The size of this manual does not provide sufficient space for showing the complete, contiguous menu structure. As a compensation a cut-up presentation of the menu structure is shown. Service menu overview /Configuration for the slave unit...
  • Page 173 MagFlux® Series EN 3.05 Version 2102 Service menu overview /Configuration for the slave unit...
  • Page 174 MagFlux® Series EN 3.05 Version 2102 Service menu overview /Configuration for the slave unit...
  • Page 175 MagFlux® Series EN 3.05 Version 2102  Menu items under Zero point calibration and Verification are for future usage only Service menu overview /Configuration for the slave unit...
  • Page 176: Display Setup Menu Overview

    MagFlux® Series EN 3.05 Version 2102 Display setup menu overview The size of this manual does not provide sufficient space for showing the complete, contiguous menu structure. As a compensation a cut-up presentation of the menu structure is shown. Display setup menu overview /Configuration for the slave unit...
  • Page 177 MagFlux® Series EN 3.05 Version 2102 Display setup menu overview /Configuration for the slave unit...
  • Page 178 In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation. For more information on how Xylem can help you, go towww.xyleminc.com MJK Automation ApS Blokken 9 DK-3460 Birkerød...